Increasing shear force microscopy scanning rate using active quality-factor control

被引:25
作者
Antognozzi, M
Szczelkun, MD
Humphris, ADL
Miles, MJ
机构
[1] Hannover Med Sch, D-30625 Hannover, Germany
[2] Univ Bristol, Sch Med Sci, Dept Biochem, Bristol BS8 1TD, Avon, England
[3] Univ Bristol, HH Wills Phys Lab, Dept Phys, Bristol BS8 1TL, Avon, England
关键词
D O I
10.1063/1.1571233
中图分类号
O59 [应用物理学];
学科分类号
摘要
The use of an active quality-factor control device has been demonstrated to increase the scanning rate of a shear force microscope by lowering the quality factor of the oscillating probe. Shear force microscopy (often used in combination with scanning near-field optical microscopy) requires slow scanning rates due to the long (>1 ms) time constant of the oscillating element (optical fiber or tuning fork). By electronically reducing the quality factor of the probe, it was possible to tune the time constant and correspondingly reduce the overall scanning time by a factor of 4. It was also seen that probes with high quality factors could manifest an unstable behavior; by reducing the quality factor of the probe, it was possible to increase the signal-to-noise ratio and so improve the imaging resolution. Deoxyribonucleic acid molecules deposited onto mica and imaged in air were used to demonstrate the technique. (C) 2003 American Institute of Physics.
引用
收藏
页码:2761 / 2763
页数:3
相关论文
共 22 条
[1]   FREQUENCY-MODULATION DETECTION USING HIGH-Q CANTILEVERS FOR ENHANCED FORCE MICROSCOPE SENSITIVITY [J].
ALBRECHT, TR ;
GRUTTER, P ;
HORNE, D ;
RUGAR, D .
JOURNAL OF APPLIED PHYSICS, 1991, 69 (02) :668-673
[2]  
Antognozzi M, 2002, SINGLE MOL, V3, P105, DOI 10.1002/1438-5171(200206)3:2/3<105::AID-SIMO105>3.0.CO
[3]  
2-#
[4]   Modeling of cylindrically tapered cantilevers for transverse dynamic force microscopy (TDFM) [J].
Antognozzi, M ;
Binger, DR ;
Humphris, ADL ;
James, PJ ;
Miles, MJ .
ULTRAMICROSCOPY, 2001, 86 (1-2) :223-232
[5]   A phase-locked shear-force microscope for distance regulation in near-field optical microscopy [J].
Atia, WA ;
Davis, CC .
APPLIED PHYSICS LETTERS, 1997, 70 (04) :405-407
[6]   COMBINED SHEAR FORCE AND NEAR-FIELD SCANNING OPTICAL MICROSCOPY [J].
BETZIG, E ;
FINN, PL ;
WEINER, JS .
APPLIED PHYSICS LETTERS, 1992, 60 (20) :2484-2486
[7]  
Biryukov AV, 2001, PHYS LOW-DIMENS STR, V3-4, P17
[8]   Influence of environmental conditions on shear-force distance control in near-field optical microscopy [J].
Brunner, R ;
Marti, O ;
Hollricher, O .
JOURNAL OF APPLIED PHYSICS, 1999, 86 (12) :7100-7106
[9]   Influence of the water layer on the shear force damping in near-field microscopy [J].
Davy, S ;
Spajer, M ;
Courjon, D .
APPLIED PHYSICS LETTERS, 1998, 73 (18) :2594-2596
[10]   Active quality factor control in liquids for force spectroscopy [J].
Humphris, ADL ;
Tamayo, J ;
Miles, MJ .
LANGMUIR, 2000, 16 (21) :7891-7894