Hard X-ray spectroscopy using a small-format TlBr array

被引:41
作者
Owens, A
Bavdaz, M
Brammertz, G
Gostilo, V
Haack, N
Kozorezov, A
Lisjutin, I
Peacock, A
Zatoloka, S
机构
[1] ESTEC, ESA, Sci Payloads Technol Div, SCI ST, NL-2200 AG Noordwijk, Netherlands
[2] Balt Sci Instruments, LV-1005 Riga, Latvia
[3] DESY, HASYLAB, D-22607 Hamburg, Germany
[4] Univ Lancaster, Dept Phys, Lancaster, England
关键词
compound semiconductors; TlBr; X-rays;
D O I
10.1016/S0168-9002(02)01562-0
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We report X-ray measurements on a prototype 3 x 3 T1Br pixel array, produced to assess the technological feasibility of making a Fano limited imager, which can operate near room temperature. The device was fabricated on monocrystalline material of size 2.7 x 2.7 x 1.0 mm(3). It has a pixel size of 350 x 350 mum(2) and pitch of 450 mum. Measurements were carried out on all pixels over the energy range 5.9-662 keV using radioactive sources. The leakage currents were found to be low enough to allow room-temperature operation, with typical energy resolutions of similar to20 keV FWHM at 59.95 keV under full-area illumination. At a reduced detector temperature of -30degreesC, these fell to similar to4keV FWHM. Although the spectral performance of the present array is currently impaired by material limitations, its spectral acuity was found to be greatly enhanced by the small pixel effect. Additional photon metrology was carried out at the Hamburger Synchrotron-strahlungslabor radiation facility. Under monochromatic pencil beam illumination, the measured energy resolutions at 20 keV were similar to3 keV FWHM at -30degreesC. The spatial uniformity of the array was measured using a 50 x 50 mum(2), 20 keV monoenergetic X-ray beam, raster scanned over the entire active area. The response, in terms of count rate, gain and energy resolution was found to be uniform at the few percent level, consistent with statistics. It was observed during these measurements, that the X-ray response of the pixels was unstable, showing time-dependent gain shifts indicative of polarization effects. The magnitude of the effect was proportional to the total energy deposition per unit time. Lastly, the use of T1Br arrays in nuclear medicine applications is discussed with particular emphasis on radio-guided surgical probes. Recommendations for an optimized design are given. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:359 / 369
页数:11
相关论文
共 15 条
[1]   MEDIPIX: a VLSI chip for a GaAs pixel detector for digital radiology [J].
Amendolia, SR ;
Bertolucci, E ;
Bisogni, MG ;
Bottigli, U ;
Ceccopieri, A ;
Ciocci, MA ;
Conti, M ;
Delogu, P ;
Fantacci, ME ;
Maestro, P ;
Marzulli, V ;
Pernigotti, E ;
Romeo, N ;
Rosso, V ;
Rosso, P ;
Stefanini, A ;
Stumbo, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 422 (1-3) :201-205
[2]   CHARGE-TRANSPORT IN ARRAYS OF SEMICONDUCTOR GAMMA-RAY DETECTORS [J].
BARRETT, HH ;
ESKIN, JD ;
BARBER, HB .
PHYSICAL REVIEW LETTERS, 1995, 75 (01) :156-159
[3]   Future space applications of compound semiconductor X-ray detectors [J].
Bavdaz, M ;
Peacock, A ;
Owens, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 458 (1-2) :123-131
[4]  
DUSI W, 1998, P IEEE T NUC SCI, V2, P754
[5]   Improved spectrometric characteristics of thallium bromide nuclear radiation detectors [J].
Hitomi, K ;
Murayama, T ;
Shoji, T ;
Suehiro, T ;
Hiratate, Y .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 428 (2-3) :372-378
[6]   Room temperature X- and gamma-ray detectors using thallium bromide crystals [J].
Hitomi, K ;
Muroi, O ;
Shoji, T ;
Suehiro, T ;
Hiratate, Y .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 436 (1-2) :160-164
[7]  
HOFSTADTER R, 1949, PHYS REV, V75, P1120
[8]  
JAZURRAHMAN I, 1987, NUCL INSTRUM METH A, V261, P427, DOI 10.1016/0168-9002(87)90350-0
[10]   THALLIUM BROMIDE SEMICONDUCTOR X-RAY AND GAMMA-RAY DETECTORS [J].
OLSCHNER, F ;
SHAH, KS ;
LUND, JC ;
ZHANG, J ;
DALEY, K ;
MEDRICK, S ;
SQUILLANTE, MR .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 322 (03) :504-508