A novel method to assess primary stability of press-fit acetabular cups

被引:36
作者
Crosnier, Emilie A. [1 ]
Keogh, Patrick S. [1 ]
Miles, Anthony W. [1 ]
机构
[1] Univ Bath, Dept Mech Engn, Ctr Orthopaed Biomech, Bath BA2 7AY, Avon, England
关键词
Total hip replacement; acetabular cups; press-fit fixation; initial stability; biomechanical study; POROUS-SURFACED IMPLANTS; TOTAL HIP-ARTHROPLASTY; INITIAL STABILITY; IN-VITRO; MECHANICAL-PROPERTIES; SCREW FIXATION; COMPONENTS; REPLACEMENT; DEFORMATION; MICROMOTION;
D O I
10.1177/0954411914557714
中图分类号
R318 [生物医学工程];
学科分类号
100103 [病原生物学];
摘要
Initial stability is an essential prerequisite to achieve osseointegration of press-fit acetabular cups in total hip replacements. Most in vitro methods that assess cup stability do not reproduce physiological loading conditions and use simplified acetabular models with a spherical cavity. The aim of this study was to investigate the effect of bone density and acetabular geometry on cup stability using a novel method for measuring acetabular cup micromotion. A press-fit cup was inserted into Sawbones((R)) foam blocks having different densities to simulate normal and osteoporotic bone variations and different acetabular geometries. The stability of the cup was assessed in two ways: (a) measurement of micromotion of the cup in 6 degrees of freedom under physiological loading and (b) uniaxial push-out tests. The results indicate that changes in bone substrate density and acetabular geometry affect the stability of press-fit acetabular cups. They also suggest that cups implanted into weaker, for example, osteoporotic, bone are subjected to higher levels of micromotion and are therefore more prone to loosening. The decrease in stability of the cup in the physiological model suggests that using simplified spherical cavities to model the acetabulum over-estimates the initial stability of press-fit cups. This novel testing method should provide the basis for a more representative protocol for future pre-clinical evaluation of new acetabular cup designs.
引用
收藏
页码:1126 / 1134
页数:9
相关论文
共 42 条
[1]
Adler E, 1992, J Arthroplasty, V7, P295, DOI 10.1016/0883-5403(92)90052-R
[2]
Primary stability of two uncemented acetabular components of different geometry: hemispherical or peripherally enhanced? [J].
Antoniades, G. ;
Smith, E. J. ;
Deakin, A. H. ;
Wearing, S. C. ;
Sarungi, M. .
BONE & JOINT RESEARCH, 2013, 2 (12) :264-269
[3]
Initial stability of a cementless acetabular cup design: Experimental investigation on the effect of adding fins to the rim of the cup [J].
Baleani, M ;
Fognani, R ;
Toni, A .
ARTIFICIAL ORGANS, 2001, 25 (08) :664-669
[4]
Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[5]
STEM CURVATURE AND LOAD ANGLE INFLUENCE THE INITIAL RELATIVE BONE-IMPLANT MOTION OF CEMENTLESS FEMORAL STEMS [J].
BERZINS, A ;
SUMNER, DR ;
ANDRIACCHI, TP ;
GALANTE, JO .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1993, 11 (05) :758-769
[6]
Preclinical testing of femoral hip components: An experimental investigation with four prostheses [J].
Britton, JR ;
Prendergast, PJ .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (05) :872-880
[7]
Clements JP, 2005, ORS WASH DC 20 23 FE
[8]
Clements JP, 2003, IASTED INT C BIOMECH
[9]
Cotogno G, 2012, TOTAL HIP ARTHROPLAS
[10]
Radiographic anatomic structure of the arthritic acetabulum and its influence on total hip arthroplasty [J].
Dorr, LD ;
Bechtol, CO ;
Watkins, RG ;
Wan, ZN .
JOURNAL OF ARTHROPLASTY, 2000, 15 (07) :890-900