Strut Coverage and Vessel Wall Response to Zotarolimus-Eluting and Bare-Metal Stents Implanted in Patients With ST-Segment Elevation Myocardial Infarction The OCTAMI (Optical Coherence Tomography in Acute Myocardial Infarction) Study

被引:54
作者
Guagliumi, Giulio [1 ]
Sirbu, Vasile [1 ]
Bezerra, Hiram [3 ]
Biondi-Zoccai, Giuseppe [2 ]
Fiocca, Luigi [1 ]
Musumeci, Giuseppe [1 ]
Matiashvili, Aleksandre [1 ]
Lortkipanidze, Nikoloz [1 ]
Tahara, Satoko [3 ]
Valsecchi, Orazio [1 ]
Costa, Marco [3 ]
机构
[1] Osped Riuniti Bergamo, Cardiovasc Dept, Div Cardiol, I-24128 Bergamo, Italy
[2] Univ Turin, Div Cardiol, Turin, Italy
[3] Case Western Reserve Univ, Univ Hosp, Case Med Ctr, Cleveland, OH 44106 USA
关键词
acute myocardial infarction; coronary artery disease; optical coherence tomography; stent coverage; zotarolimus-eluting stent; INTRAVASCULAR ULTRASOUND; NEOINTIMAL COVERAGE; THROMBUS; METAANALYSIS; RECOVERY; TRIALS;
D O I
10.1016/j.jcin.2010.04.005
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives Using optical coherence tomography, we assessed the proportion of uncovered struts at 6-month follow-up in zotarolimus-eluting stents (ZES), specifically Endeavor (Medtronic CardioVascular, Santa Rosa, California) stents, and identical bare-metal stents (BMS) implanted in patients with ST-segment elevation myocardial infarction (STEMI). Background Sirolimus- and paclitaxel-eluting stents implanted in STEMI have been associated with delayed healing and incomplete strut coverage. ZES are associated with a more complete and uniform strut coverage in stable patients, but whether this holds true also after STEMI is unknown. Methods Forty-four patients with STEMI who underwent primary PCI were randomized to ZES or BMS (3:1 randomization). Angiographic, intravascular ultrasound, and optical coherence tomography follow-up was conducted at 6 months and clinical follow-up at 1 year. All images were analyzed by an independent core laboratory that was blind to stent assignments. Results There were no differences between ZES and BMS in percentage of uncovered struts (median: 0.00% [interquartile range (IQR): 0.00% to 1.78%] vs. 1.98% [IQR: 0.21% to 7.33%], p = 0.13), maximum length of uncovered segments (0.00 [IQR: 0.00 to 1.19] mm vs. 1.38 [IQR: 0.65 to 3.30] mm, p = 0.10), percentage of malapposed struts (0.00% [IQR: 0.00% to 0.23%] vs. 0.15% [IQR: 0.00% to 5.81%], p = 0.16), and maximum length of malapposed segments (0.00 [IQR: 0.00 to 0.67] mm vs. 0.33 [IQR: 0.00 to 2.55] mm, p = 0.20). Neointimal response was similar between ZES and BMS (332 [IQR: 240 to 429] mu m vs. 186 [IQR: 136 to 348] mu m, p = 0.99) and evenly distributed. No late acquired malapposition was observed in both groups. There were no deaths, myocardial infarction, or stent thromboses at 1 year. Conclusions This optical coherence tomography study found no difference in strut coverage and similar vessel response to ZES, when compared with identical BMS, implanted during primary percutaneous coronary intervention in STEMI patients. (Six-Month Coverage and Vessel Wall Response of the Zotarolimus Drug-Eluting Stent Implanted in AMI Assessed by Optical Coherence Tomography [OCTAMI]; NCT00704561) (J Am Coll Cardiol Intv 2010;3:680-7) (C) 2010 by the American College of Cardiology Foundation
引用
收藏
页码:680 / 687
页数:8
相关论文
共 30 条
[1]   Intracoronary Optical Coherence Tomography: A Comprehensive Review Clinical and Research Applications [J].
Bezerra, Hiram G. ;
Costa, Marco A. ;
Guagliumi, Giulio ;
Rollins, Andrew M. ;
Simon, Daniel I. .
JACC-CARDIOVASCULAR INTERVENTIONS, 2009, 2 (11) :1035-1046
[2]   Correlation of Intravascular Ultrasound Findings With Histopathological Analysis of Thrombus Aspirates in Patients With Very Late Drug-Eluting Stent Thrombosis [J].
Cook, Stephane ;
Ladich, Elena ;
Nakazawa, Gaku ;
Eshtehardi, Parham ;
Neidhart, Michel ;
Vogel, Rolf ;
Togni, Mario ;
Wenaweser, Peter ;
Billinger, Michael ;
Seiler, Christian ;
Gay, Steffen ;
Meier, Bernhard ;
Pichler, Werner J. ;
Jueni, Peter ;
Virmani, Renu ;
Windecker, Stephan .
CIRCULATION, 2009, 120 (05) :391-399
[3]   Clinical end points in coronary stent trials - A case for standardized definitions [J].
Cutlip, Donald E. ;
Windecker, Stephan ;
Mehran, Roxana ;
Boam, Ashley ;
Cohen, David J. ;
van Es, Gerrit-Anne ;
Steg, P. Gabriel ;
Morel, Marie-angele ;
Mauri, Laura ;
Vranckx, Pascal ;
McFadden, Eugene ;
Lansky, Alexandra ;
Hamon, Martial ;
Krucoff, Mitchell W. ;
Serruys, Patrick W. .
CIRCULATION, 2007, 115 (17) :2344-2351
[4]   Short and long-term benefits of sirolimus-eluting stent in ST-segment elevation myocardial infarction: a meta-analysis of randomized trials [J].
De Luca, Giuseppe ;
Valgimigli, Marco ;
Spaulding, Christian ;
Menichelli, Maurizio ;
Brunner-La Rocca, Hans Peter ;
van der Hoeven, Bas L. ;
Di Lorenzo, Emilio ;
de la Llera, Luis-S. Diaz ;
Pasceri, Vincenzo ;
Pittl, Undine ;
Percoco, Gianfranco ;
Violini, Roberto ;
Stone, Gregg W. .
JOURNAL OF THROMBOSIS AND THROMBOLYSIS, 2009, 28 (02) :200-210
[5]   Vascular responses to drug eluting stents - Importance of delayed healing [J].
Finn, Aloke V. ;
Nakazawa, Gaku ;
Joner, Michael ;
Kolodgie, Frank D. ;
Mont, Erik K. ;
Gold, Herman K. ;
Virmani, Renu .
ARTERIOSCLEROSIS THROMBOSIS AND VASCULAR BIOLOGY, 2007, 27 (07) :1500-1510
[6]   Reproducibility of quantitative optical coherence tomography for stent analysis [J].
Gonzalo, Nieves ;
Garcia-Garcia, Hector M. ;
Serruys, Patrick W. ;
Commissaris, Koen H. ;
Bezerra, Hiram ;
Gobbens, Pierre ;
Costa, Marco ;
Regar, Evelyn .
EUROINTERVENTION, 2009, 5 (02) :224-232
[7]   Incomplete Stent Apposition and Delayed Tissue Coverage Are More Frequent in Drug-Eluting Stents Implanted During Primary Percutaneous Coronary Intervention for ST-Segment Elevation Myocardial Infarction Than in Drug-Eluting Stents Implanted for Stable/Unstable Angina Insights From Optical Coherence Tomography [J].
Gonzalo, Nieves ;
Barlis, Peter ;
Serruys, Patrick W. ;
Garcia-Garcia, Hector M. ;
Onuma, Yoshinobu ;
Ligthart, Jurgen ;
Regar, Evelyn .
JACC-CARDIOVASCULAR INTERVENTIONS, 2009, 2 (05) :445-452
[8]  
GUAGLIUMI G, 2008, CIRCULATION, V118, P2315
[9]   Optical coherence tomography: High resolution intravascular imaging to evaluate vascular healing after coronary stenting [J].
Guagliumi, Giulio ;
Sirbu, Vasile .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2008, 72 (02) :237-247
[10]   Thrombosis modulates arterial drug distribution for drug-eluting stents [J].
Hwang, CW ;
Levin, AD ;
Jonas, M ;
Li, PH ;
Edelman, ER .
CIRCULATION, 2005, 111 (13) :1619-1626