Centrifugal microfluidics for biomedical applications

被引:587
作者
Gorkin, Robert [2 ]
Park, Jiwoon [1 ]
Siegrist, Jonathan [2 ]
Amasia, Mary [2 ]
Lee, Beom Seok [3 ]
Park, Jong-Myeon [3 ,4 ]
Kim, Jintae [3 ]
Kim, Hanshin [3 ]
Madou, Marc [1 ,2 ]
Cho, Yoon-Kyoung [1 ]
机构
[1] UNIST, Ulsan 689798, South Korea
[2] Univ Calif Irvine, Irvine, CA 92697 USA
[3] Samsung Elect Co Ltd, Suwon 443746, Gyeonggi Do, South Korea
[4] Yonsei Univ, Seoul 120749, South Korea
关键词
WHOLE-BLOOD; COMPACT DISC; CELL-LYSIS; DNA; IMMUNOASSAYS; PLATFORM; SYSTEM; DEVICE; FLOW; CD;
D O I
10.1039/b924109d
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The centrifugal microfluidic platform has been a focus of academic and industrial research efforts for almost 40 years. Primarily targeting biomedical applications, a range of assays have been adapted on the system; however, the platform has found limited commercial success as a research or clinical tool. Nonetheless, new developments in centrifugal microfluidic technologies have the potential to establish wide-spread utilization of the platform. This paper presents an in-depth review of the centrifugal microfluidic platform, while highlighting recent progress in the field and outlining the potential for future applications. An overview of centrifugal microfluidic technologies is presented, including descriptions of advantages of the platform as a microfluidic handling system and the principles behind centrifugal fluidic manipulation. The paper also discusses a history of significant centrifugal microfluidic platform developments with an explanation of the evolution of the platform as it pertains to academia and industry. Lastly, we review the few centrifugal microfluidic-based sample-to-answer analysis systems shown to date and examine the challenges to be tackled before the centrifugal platform can be more broadly accepted as a new diagnostic platform. In particular, fully integrated, easy to operate, inexpensive and accurate microfluidic tools in the area of in vitro nucleic acid diagnostics are discussed.
引用
收藏
页码:1758 / 1773
页数:16
相关论文
共 69 条
[51]   Read-out concepts for multiplexed bead-based fluorescence immunoassays on centrifugal microfluidic platforms [J].
Riegger, L ;
Grumann, M ;
Nann, T ;
Riegler, J ;
Ehlert, O ;
Bessler, W ;
Mittenbuehler, K ;
Urban, G ;
Pastewka, L ;
Brenner, T ;
Zengerle, R ;
Ducrée, J .
SENSORS AND ACTUATORS A-PHYSICAL, 2006, 126 (02) :455-462
[52]  
Rocco R.M., 2006, Landmark papers in clinical chemistry
[53]  
SCHEMBRI CT, 1992, CLIN CHEM, V38, P1665
[54]   CENTRIFUGATION AND CAPILLARITY INTEGRATED INTO A MULTIPLE ANALYTE WHOLE-BLOOD ANALYZER [J].
SCHEMBRI, CT ;
BURD, TL ;
KOPFSILL, AR ;
SHEA, LR ;
BRAYNIN, B .
JOURNAL OF AUTOMATIC CHEMISTRY, 1995, 17 (03) :99-104
[55]   A continuous-flow capillary mixing method to monitor reactions on the microsecond time scale [J].
Shastry, MCR ;
Luck, SD ;
Roder, H .
BIOPHYSICAL JOURNAL, 1998, 74 (05) :2714-2721
[56]  
Shea M., 2003, JALA, V8, P74
[57]   Validation of a centrifugal microfluidic sample lysis and homogenization platform for nucleic acid extraction with clinical samples [J].
Siegrist, Jonathan ;
Gorkin, Robert ;
Bastien, Martine ;
Stewart, Gale ;
Peytavi, Regis ;
Kido, Horacio ;
Bergeron, Michel ;
Madou, Marc .
LAB ON A CHIP, 2010, 10 (03) :363-371
[58]   Direct hemoglobin measurement on a centrifugal microfluidic platform for point-of-care diagnostics [J].
Steigert, J. ;
Grumann, M. ;
Dube, M. ;
Streule, W. ;
Riegger, L. ;
Brenner, T. ;
Koltay, P. ;
Mittmann, K. ;
Zengerle, R. ;
Ducree, J. .
SENSORS AND ACTUATORS A-PHYSICAL, 2006, 130 (SPEC. ISS.) :228-233
[59]   Fully integrated whole blood testing by real-time absorption measurement on a centrifugal platform [J].
Steigert, J. ;
Grumann, M. ;
Brenner, T. ;
Riegger, L. ;
Harter, J. ;
Zengerle, R. ;
Ducree, J. .
LAB ON A CHIP, 2006, 6 (08) :1040-1044
[60]  
Steigert J., 2005, JALA J ASS LAB AUTOM, V10, P331