Intramyocardial pacing and sensing for the enhancement of cardiac stimulation and sensing specificity

被引:12
作者
Asirvatham, Samuel J. [1 ]
Bruce, Charles J. [1 ]
Danielsen, Andrew [1 ]
Johnson, Susan B. [1 ]
Okumura, Yasuo [1 ]
Kathmann, Eva [1 ]
Packer, Douglas L. [1 ]
Friedman, Paul A. [1 ]
机构
[1] Mayo Clin, Div Cardiovasc Med, Rochester, MN 55905 USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2007年 / 30卷 / 06期
关键词
new technology; pacing;
D O I
10.1111/j.1540-8159.2007.00745.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Intracardiac electrodes create an "antenna" capable of unintentionally recording and stimulating tissue beyond the chamber in which they are positioned, resulting in far-field R wave oversensing in pacemakers and inappropriate detection in defibrillators. This feasibility study sought to determine whether a specially constructed lead with two distal totally intramyocardial electrodes could overcome these limitations. Methods: Two mongrel dogs were anesthetized and a median sternotomy performed. Epicardial intramyocardial pacing and sensing function was assessed and compared to standard active fixation pacing and sensing placed at the same atrial and ventricular sites. Right ventricular pacing was also assessed. Results: For the novel intramyocardial lead, the average R wave amplitude was 7.2 mV, compared to an average R wave of 8.4 mV for the standard active fixation lead placed at identical ventricular sites; P-waves were also similar. Cross-chamber sensing was present in the ventricle and atrium with the standard lead, and absent with the intramyocardial lead. The average pacing threshold was 0.7 MA at 0.2 ins for the novel lead compared to 1.1 mA for the standard lead. With the standard lead, phrenic stimulation was seen at threshold (cathode distal) and at 3 mA (cathode proximal electrode). No phrenic stimulation was seen with the novel intramyocardial lead despite outputs up to 20 mA at sites located 3-5 mm from the phrenic nerve. Conclusion: Totally intramyocardial pacing is feasible, and results in site-specific pacing and sensing function. This may eliminate far-field signal oversensing and phrenic stimulation in future devices.
引用
收藏
页码:748 / 754
页数:7
相关论文
共 17 条
[1]   Far-field R-Wave reduction with a novel lead design: Experimental and human results [J].
De Voogt, W ;
Van Hemel, N ;
Willems, A ;
Visser, J ;
Chitre, Y ;
Bornzin, G ;
Helland, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (08) :782-788
[2]   Direct his-bundle pacing: Present and future [J].
Deshmukh, PM ;
Romanyshyn, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (06) :862-870
[3]   Dual-chamber versus single-chamber detection enhancements for implantable defibrillator rhythm diagnosis - The detect supraventricular tachycardia study [J].
Friedman, PA ;
McClelland, RL ;
Bamlet, WR ;
Acosta, H ;
Kessler, D ;
Munger, TM ;
Kavesh, NG ;
Wood, M ;
Daoud, E ;
Massumi, A ;
Schuger, C ;
Shorofsky, S ;
Wilkoff, B ;
Glikson, M .
CIRCULATION, 2006, 113 (25) :2871-2879
[4]   Randomized prospective pilot study of long-term dual-site atrial pacing for prevention of atrial fibrillation [J].
Friedman, PA ;
Hill, MRS ;
Hammill, SC ;
Hayes, DL ;
Stanton, MS .
MAYO CLINIC PROCEEDINGS, 1998, 73 (09) :848-854
[5]   ENDOCARDIAL THERESHOLD OF CARDIAC RESPONSE AS A FUNCTION OF ELECTRODE SURFACE AREA [J].
FURMAN, S ;
PARKER, B ;
ESCHER, DJW ;
SOLOMON, N .
JOURNAL OF SURGICAL RESEARCH, 1968, 8 (04) :161-&
[6]  
FURMAN S, 1971, J SURG RES, P11105
[7]   FAILURE OF DEMAND PACING WITH SMALL SURFACE-AREA ELECTRODES [J].
HUGHES, HC ;
BROWNLEE, RR ;
TYERS, GFO .
CIRCULATION, 1976, 54 (01) :128-132
[8]   Far-field R wave oversensing in dual chamber pacemakers designed for atrial arrhythmia management: Effect of pacing site and lead tip to ring distance [J].
Inama, G ;
Santini, M ;
Padeletti, L ;
Boriani, G ;
Botto, G ;
Capucci, A ;
Gulizia, M ;
Ricci, R ;
Rizzon, P ;
Ferri, F ;
Miraglia, F ;
Raneri, R ;
Grammatico, A .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (09) :1221-1230
[9]   Severe symptomatic tricuspid valve regurgitation due to permanent pacemaker or implantable cardioverter-defibrillator leads [J].
Lin, G ;
Nishimura, RA ;
Connolly, HM ;
Dearani, JA ;
Sundt, TM ;
Hayes, DL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 45 (10) :1672-1675
[10]   Inhibition of biventricular pacemakers by oversensing of far-field atrial depolarization [J].
Lipchenca, I ;
Garrigue, S ;
Glikson, M ;
Barold, SS ;
Clémenty, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (03) :365-367