Transvenous defibrillation leads: High incidence of failure during long-term follow-up

被引:85
作者
Dorwarth, U
Frey, B
Dugas, M
Matis, T
Fiek, M
Schmoeckel, M
Remp, T
Durchlaub, I
Gerth, A
Steinbeck, G
Hoffmann, E
机构
[1] Univ Munich, Dept Med 1, D-80539 Munich, Germany
[2] Univ Munich, Dept Med Informat, D-80539 Munich, Germany
[3] Univ Munich, Dept Cardiac Surg, D-80539 Munich, Germany
关键词
implantable cardioverter defibrillator; transvenous leads; insulation failure; inappropriate shocks; polyurethane insulation;
D O I
10.1046/j.1540-8167.2003.02305.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Patients with implantable cardioverter defibrillators (ICD) critically depend on correct functioning of their system. The aim of this study was to determine the incidence and clinical presentation of transvenous ICD lead failures during long-term follow-up. Methods and Results: The study group consisted of 261 consecutive patients who received Medtronic right ventricular polyurethane transvenous leads (models 6884, 6966, 6936) between 1990 and 1998 as part of an abdominal (n = 70) or pectoral (n = 191) ICD system. During mean follow-up of 4.0 +/- 2.6 years, 31 patients (12%) developed a lead-related sensing failure with oversensing of artifacts. All failures except two were compatible with an insulation defect and occurred late after ICD placement (6.0 +/- 1.8 years after implant). Lead survival decreased from 98% at 4-year follow-up to only 62% at 8-year follow-up. Lead survival was not related to patient age, sex, venous lead implantation route, or device implantation site. In 26 (87%) of 31 patients, a sensing defect resulted in inappropriate detection of ventricular fibrillation and subsequent delivery of 3 +/- 3 (range 1-11) inappropriate shocks in 19 (61%) of 31 patients. Device interrogation showed artifacts classified as nonsustained ventricular tachycardia in 21 patients, 40 +/- 43 days before clinically relevant failure of the system. One patient with a subclavian crush syndrome required resuscitation because of undersensing of true ventricular fibrillation. Conclusion: Transvenous polyurethane ICD leads showed a high rate of lead insulation failure late after implantation with frequent inappropriate shock deliveries. Close follow-up is mandatory in patients with these leads. Automated device control features with patient alert function integrated into new devices may contribute to early detection of lead failure.
引用
收藏
页码:38 / 43
页数:6
相关论文
共 20 条
  • [1] Insulation lead failure: Is it a matter of insulation coating, venous approach, or both?
    Antonelli, D
    Rosenfeld, T
    Freedberg, NA
    Palma, E
    Gross, JN
    Furman, S
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (02): : 418 - 421
  • [2] Electrophysiologic testing to identify patients with coronary artery disease who are at risk for sudden death
    Buxton, AE
    Lee, KL
    DiCarlo, L
    Gold, MR
    Greer, GS
    Prystowsky, EN
    O'Toole, MF
    Tang, A
    Fisher, JD
    Coromilas, J
    Talajic, M
    Hafley, G
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2000, 342 (26) : 1937 - 1945
  • [3] Incidence and implications of abrasion of implantable cardioverter-defibrillator leads
    DeLurgio, DB
    Sathavorn, C
    Mera, F
    Leon, A
    Walter, PF
    Langberg, JJ
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1997, 79 (10) : 1409 - &
  • [4] Lead fracture in cephalic versus subclavian approach with transvenous implantable cardioverter defibrillator systems
    Gallik, DM
    BenZur, UM
    Gross, JN
    Furman, S
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (07): : 1089 - 1094
  • [5] Feasibility and initial results of an Internet-based pacemaker and ICD pulse generator and lead registry
    Hauser, R
    Hayes, D
    Parsonnet, V
    Furman, S
    Epstein, A
    Hayes, J
    Saksena, S
    Irwin, M
    Almquist, A
    Cannom, D
    Gross, J
    Kallinen, L
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (01): : 82 - 87
  • [6] MULTICENTER EXPERIENCE WITH A BIPOLAR TINED POLYURETHANE VENTRICULAR LEAD
    HAYES, DL
    GRAHAM, KJ
    IRWIN, M
    VIDAILLET, H
    DISLER, G
    SWEESY, M
    KINCAID, D
    OSBORN, MJ
    SUMAN, VJ
    NEUBAUER, SA
    SEEBANDT, M
    KALLINEN, L
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (05): : 999 - 1004
  • [7] LONG-TERM PERFORMANCE OF ENDOCARDIAL PACING LEADS
    HELGUERA, ME
    MALONEY, JD
    PINSKI, SL
    WOSCOBOINIK, JR
    WILKOFF, BL
    CASTLE, LW
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (01): : 56 - 64
  • [8] INCIDENCE OF ICD LEAD RELATED COMPLICATIONS DURING LONG-TERM FOLLOW-UP - COMPARISON OF EPICARDIAL AND ENDOCARDIAL ELECTRODE SYSTEMS
    KORTE, T
    JUNG, W
    SPEHL, S
    WOLPERT, C
    MOOSDORF, R
    MANZ, M
    LUDERITZ, B
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (11): : 2053 - 2061
  • [9] Should lead explantation be the practice standard when a lead needs to be replaced?
    Levine, PA
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (04): : 421 - 422
  • [10] McAnulty J, 1997, NEW ENGL J MED, V337, P1576