Accuracy of an image-free cup navigation system -: an anatomical study

被引:18
作者
Kalteis, T
Beckmann, J
Herold, T
Zysk, S
Bäthis, H
Perlick, L
Grifka, J
机构
[1] Univ Regensburg, Orthopad Klin, Bayerischen Rheuma & Orthopad Zentrum, gGmbH BRK, D-93077 Bad Abbach, Germany
[2] Univ Regensburg, Inst Rontgendiagnost, D-8400 Regensburg, Germany
[3] Univ Munich, Orthopad Klin, Munich, Germany
来源
BIOMEDIZINISCHE TECHNIK | 2004年 / 49卷 / 09期
关键词
computer-assisted surgery; image-free navigation; hip arthroplasty; hip endoprosthesis;
D O I
10.1515/BMT.2004.048
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
The position of the acetabular cup is of decisive importance for. the function of a total hip replacement (THR). Using the conventional surgical technique, correct placement of the cup often fails due to a lack of information about pelvic tilt. With CT-based and fluoroscopically-assisted navigation procedures the accuracy of implantation has been significantly improved. However, additional radiation exposure, high cost and the increased time requirement have hampered the acceptance of these techniques. The present anatomical study evaluates the accuracy of an alternative procedure-image-free navigation. This method requires little extra effort, does not substantially delay surgery, and needs no additional imaging. Press-fit cups were implanted in 10 human cadaveric hips with the help of the image-free navigation system, and the position of the cups was checked intraoperatively with a CT-based navigation system and postoperatively by computed tomography. All cups were implanted within the targeted safe zone with an average inclination of 44degrees (range 40degrees-48degrees, SABW 2.7degrees) and an average anteversion of 18degrees (range 12-24degrees, SABW 4.1degrees). Analysis of accuracy of the image-free navigation software revealed only a small, clinically tolerable deviation in cup anteversion and cup inclination in comparison with the CT-based navigation system and the post operative CT scans. The evaluated image-free navigation system appears to be a practicable and reliable alternative to the computer-assisted implantation of acetabular cups in total hip arthroplasty.
引用
收藏
页码:257 / 262
页数:6
相关论文
共 20 条
[1]
Bernsmann K, 2001, Z ORTHOP GRENZGEB, V139, P512
[2]
Effects of acetabular abduction on cup wear rates in total hip arthroplasty [J].
Del Schutte, H ;
Lipman, AJ ;
Bannar, SM ;
Livermore, JT ;
Ilstrup, D ;
Morrey, BF .
JOURNAL OF ARTHROPLASTY, 1998, 13 (06) :621-626
[3]
DeWal Hargovind, 2003, Am J Orthop (Belle Mead NJ), V32, P377
[4]
Comparison of a mechanical acetabular alignment guide with computer placement of the socket [J].
DiGioia, AM ;
Jaramaz, B ;
Plakseychuk, AY ;
Moody, JE ;
Nikou, C ;
LaBarca, RS ;
Levison, TJ ;
Picard, F .
JOURNAL OF ARTHROPLASTY, 2002, 17 (03) :359-364
[5]
Mini-incision technique for total hip arthroplasty with navigation [J].
DiGioia, AM ;
Plakseychuk, AY ;
Levison, TJ ;
Jaramaz, B .
JOURNAL OF ARTHROPLASTY, 2003, 18 (02) :123-128
[6]
Computer-assisted navigated cup placement in primary and secondary dysplastic hips [J].
Haaker, R ;
Tiedjen, K ;
Rubenthaler, F ;
Stockheim, M .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 2003, 141 (01) :105-111
[7]
Accuracy of intraoperative assessment of acetabular prosthesis placement [J].
Hassan, DM ;
Johnston, GHF ;
Dust, WNC ;
Watson, G ;
Dolovich, AT .
JOURNAL OF ARTHROPLASTY, 1998, 13 (01) :80-84
[8]
Hirakawa K, 2001, CLIN ORTHOP RELAT R, P135
[9]
Computer assisted measurement of cup placement in total hip replacement [J].
Jaramaz, B ;
DiGioia, AM ;
Blackwell, M ;
Nikou, C .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1998, (354) :70-81
[10]
Jaramaz B, 1998, CLIN ORTHOP RELAT R, V355, P8