Simultaneous analysis of metabisulfite and sulfate by CE with indirect UV detection. Application to and validation for a pharmaceutical formulation

被引:16
作者
Geiser, L [1 ]
Varesio, E [1 ]
Veuthey, JL [1 ]
机构
[1] Univ Geneva, Lab Pharmaceut Analyt Chem, Sch Pharm, CH-1211 Geneva 4, Switzerland
关键词
capillary electrophoresis; indirect UV; inorganic anions; metabisulfite; sulfate; validation;
D O I
10.1016/S0731-7085(02)00651-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Metabisulfite is used as an antioxidant agent in a number of pharmaceutical formulations. In order to quantify simultaneously both metabisulfite and its oxidation product (sulfate), a capillary zone electrophoretic (CZE) method with indirect UV detection was developed. Best results were achieved with a background electrolyte (BGE) constituted of 15 mM pyromellitic acid, 15 mM tris-(hydroxymethyl)-aminomethane and 0.2 mM tetradecyltrimethylammonium bromide at pH 8.3 and an applied electrical field of 123 V/cm in a 32.5 cm fused silica capillary. Indirect UV detection was performed at a wavelength of 225 nm. In order to validate this method, an internal standard (IS), namely ammonium formate, was used. Moreover, due to the high chloride concentration in the pharmaceutical formulation, conductivity was adjusted by adding sodium chloride into standard solutions to prevent matrix effect. Linearity and accuracy were successfully tested in a concentration range of 33.3-250 mug/ml for sodium metabisulfite and of 50-375 mug/ml for sodium sulfate. Method precision was determined on six samples each day. Thereby, relative standard deviations (R.S.D.) of 6% and 12-13% were obtained for intra-day and inter-day precision, respectively. Considering the instability of metabisulfite and its use as an antioxidant agent and not as an active principle, the method was accepted and used for routine analyses. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1059 / 1064
页数:6
相关论文
共 21 条
[1]  
[Anonymous], STP PHARM PRATIQUES
[2]   Separation of chlorine-containing anions by ion chromatography and capillary electrophoresis [J].
Biesaga, M ;
Kwiatkowska, M ;
Trojanowicz, M .
JOURNAL OF CHROMATOGRAPHY A, 1997, 777 (02) :375-381
[3]   EVALUATION OF CARRIER ELECTROLYTES FOR CAPILLARY ZONE ELECTROPHORESIS OF LOW-MOLECULAR-MASS ANIONS WITH INDIRECT UV DETECTION [J].
COUSINS, SM ;
HADDAD, PR ;
BUCHBERGER, W .
JOURNAL OF CHROMATOGRAPHY A, 1994, 671 (1-2) :397-402
[4]   Factors influencing the choice of buffer in background electrolytes for indirect detection of fast anions by capillary electrophoresis [J].
Doble, P ;
Macka, M ;
Haddad, PR .
ELECTROPHORESIS, 1998, 19 (12) :2257-2261
[5]   Recent developments in the separation of inorganic and small organic ions by capillary electrophoresis [J].
Fritz, JS .
JOURNAL OF CHROMATOGRAPHY A, 2000, 884 (1-2) :261-275
[6]   Comparison of different electroosmotic flow modifiers in the analysis of inorganic anions by capillary electrophoresis [J].
Galceran, MT ;
Puignou, L ;
Diez, M .
JOURNAL OF CHROMATOGRAPHY A, 1996, 732 (01) :167-174
[7]   Developments in sample preparation and separation techniques for the determination of inorganic ions by ion chromatography and capillary electrophoresis [J].
Haddad, PR ;
Doble, P ;
Macka, M .
JOURNAL OF CHROMATOGRAPHY A, 1999, 856 (1-2) :145-177
[9]   PARAMETERS INFLUENCING SEPARATION AND DETECTION OF ANIONS BY CAPILLARY ELECTROPHORESIS [J].
HARROLD, MP ;
WOJTUSIK, MJ ;
RIVIELLO, J ;
HENSON, P .
JOURNAL OF CHROMATOGRAPHY, 1993, 640 (1-2) :463-471
[10]   OPTIMIZATION OF DETECTION SENSITIVITY IN THE CAPILLARY ELECTROPHORESIS OF INORGANIC ANIONS [J].
JANDIK, P ;
JONES, WR .
JOURNAL OF CHROMATOGRAPHY, 1991, 546 (1-2) :431-443