Ceramic-based multisite microelectrode array for rapid choline measures in brain tissue

被引:56
作者
Burmeister, JJ
Palmer, M
Gerhardt, GA [1 ]
机构
[1] Univ Kentucky, Dept Anat,Ctr Sensor Technol, Morris K Udall Parkinsons Dis Res Ctr Excellence, Chandler Med Ctr, Lexington, KY 40536 USA
[2] Univ Kentucky, Dept Neurobiol & Neurol,Ctr Sensor Technol, Morris K Udall Parkinsons Dis Res Ctr Excellence, Chandler Med Ctr, Lexington, KY 40536 USA
[3] Univ Colorado, Hlth Sci Ctr, Dept Pharmacol, Denver, CO 80202 USA
基金
美国国家科学基金会;
关键词
choline; neurotransmitters; voltammetry; biosensor; CNS;
D O I
10.1016/S0003-2670(03)00067-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A ceramic-based multisite microelectrode array to measure choline in vivo in brain tissues is described. The microelectrodes were linear to 200 muM choline (R-2 = 0.999 +/- 0.001) with a detection limit of approximately 0.4 muM (S/N of 3) in both single microelectrode and self-referencing amperometric recording modes. The 90% rise time of the sensor was 1.4 s, allowing for rapid measures of choline. Good selectivity (>300:1) was observed over interferents such as ascorbic acid, uric acid, and DOPAC in the single microelectrode mode. However, a self-referencing recording mode was needed to remove potassium-evoked dopamine signals in rat striatum. In vivo measurements of choline in the rat brain are presented. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:65 / 74
页数:10
相关论文
共 43 条
[1]   ANTICHOLINESTERASE ACTIVITY MEASUREMENT BY A CHOLINE BIOSENSOR - APPLICATION IN WATER ANALYSIS [J].
BERNABEI, M ;
CHIAVARINI, S ;
CREMISINI, C ;
PALLESCHI, G .
BIOSENSORS & BIOELECTRONICS, 1993, 8 (05) :265-271
[2]   Improved ceramic-based multisite microelectrode for rapid measurements of L-glutamate in the CNS [J].
Burmeister, JJ ;
Pomerleau, F ;
Palmer, M ;
Day, BK ;
Huettl, P ;
Gerhardt, GA .
JOURNAL OF NEUROSCIENCE METHODS, 2002, 119 (02) :163-171
[3]   Self referencing ceramic based multisite microelectrodes for the detection and elimination of interferences from the measurement of L-glutamate and other analytes [J].
Burmeister, JJ ;
Gerhardt, GA .
ANALYTICAL CHEMISTRY, 2001, 73 (05) :1037-1042
[4]   Ceramic-based multisite microelectrodes for electrochemical recordings [J].
Burmeister, JJ ;
Moxon, K ;
Gerhardt, GA .
ANALYTICAL CHEMISTRY, 2000, 72 (01) :187-192
[5]  
BURMEISTER JJ, UNPUB
[6]   DETERMINATION OF COMPOUNDS WITH ANTICHOLINESTERASE ACTIVITY IN COMMERCIAL DRUGS BY A NEW ENZYME SENSOR [J].
CAMPANELLA, L ;
COCCO, R ;
TOMASSETTI, M .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1992, 10 (10-12) :741-749
[7]   THE DETERMINATION OF CHOLINE IN PHARMACEUTICAL PREPARATIONS BY MEANS OF AN ENZYME SENSOR [J].
CAMPANELLA, L ;
TOMASSETTI, M ;
SAMMARTINO, MP ;
STEFANONI, P .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1989, 7 (06) :765-770
[8]   CLEARANCE OF EXOGENOUS DOPAMINE IN RAT DORSAL STRIATUM AND NUCLEUS-ACCUMBENS - ROLE OF METABOLISM AND EFFECTS OF LOCALLY APPLIED UPTAKE INHIBITORS [J].
CASS, WA ;
ZAHNISER, NR ;
FLACH, KA ;
GERHARDT, GA .
JOURNAL OF NEUROCHEMISTRY, 1993, 61 (06) :2269-2278
[9]  
COOPER J, 1996, BIOCH BASIS NEUROPHA
[10]   Application of in vivo microdialysis to the study of cholinergic systems [J].
Day, JC ;
Kornecook, TJ ;
Quirion, R .
METHODS, 2001, 23 (01) :21-39