A piezoelectric quartz crystal sensor for the determination of coagulation time in plasma and whole blood

被引:45
作者
Cheng, TJ
Chang, HC [1 ]
Lin, TM
机构
[1] Natl Cheng Kung Univ, Inst Biomed Engn, Tainan 70101, Taiwan
[2] Natl Cheng Kung Univ, Dept Med Technol, Tainan 70101, Taiwan
关键词
piezoelectric quartz crystal; blood coagulation; prothrombin time; clotting time; recalcification time;
D O I
10.1016/S0956-5663(97)00111-5
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A Piezoelectric Quartz Crystal (PQC) system possessing an adequate sensitivity to a slight viscosity change is employed to monitor the blood coagulation. In the case of an anticoagulated plasma specimen, prothrombin time (PT) was determined by the PQC sensor on the basis of the introduction of tissue thromboplastin (TF) and calcium ion. Discussing the PQC data statistically, we found that at 90% of the total frequency shift (Delta f) the best linear relationship in a double-logarithmic plot of PT versus TF concentration in the range of 5.466 similar to 22.311 unit/ml is shown. The PQC result was compared with commercial optical coagulometry and showed a good linear correlation (ca. 0.98). The clotting and recalcification of whole blood were also studied by the PQC system. The testing time for one measurement could be completed within 20 min. and had a good agreement with that obtained conventionally. (C) 1998 Elsevier Science S.A. All rights reserved.
引用
收藏
页码:147 / 156
页数:10
相关论文
共 12 条
  • [1] LECTIN COATED PIEZOELECTRIC CRYSTAL BIOSENSORS
    BARNES, C
    DSILVA, C
    JONES, JP
    LEWIS, TJ
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1992, 7 (1-3) : 347 - 350
  • [2] Brown B.A., 1988, HAEMATOLOGY PRINCIPL, P195
  • [3] KANAZAWA KK, 1985, ANAL CHIM ACTA, V175, P99
  • [4] A COATED PIEZOELECTRIC CRYSTAL SENSOR FOR ACETIC-ACID VAPOR DETERMINATION
    LEYVA, JAM
    DECISNEROS, JLHH
    DEBARREDA, DGG
    [J]. TALANTA, 1993, 40 (11) : 1725 - 1729
  • [5] A QUARTZ CRYSTAL VISCOSITY SENSOR FOR MONITORING COAGULATION REACTION AND ITS APPLICATION TO A MULTICHANNEL COAGULATION DETECTOR
    MURAMATSU, H
    KIMURA, K
    ATAKA, T
    HOMMA, R
    MIURA, Y
    KARUBE, I
    [J]. BIOSENSORS & BIOELECTRONICS, 1991, 6 (04) : 353 - 358
  • [6] QUARTZ-CRYSTAL GELATION DETECTOR FOR THE DETERMINATION OF FIBRINOGEN CONCENTRATION
    MURAMATSU, H
    TAMIYA, E
    SUZUKI, M
    KARUBE, I
    [J]. ANALYTICA CHIMICA ACTA, 1989, 217 (02) : 321 - 326
  • [7] QUARTZ-CRYSTAL MICROBALANCE FOR THE DETECTION OF MICROGRAM QUANTITIES OF HUMAN SERUM-ALBUMIN - RELATIONSHIP BETWEEN THE FREQUENCY CHANGE AND THE MASS OF PROTEIN ADSORBED
    MURATSUGU, M
    OHTA, F
    MIYA, Y
    HOSOKAWA, T
    KUROSAWA, S
    KAMO, N
    IKEDA, H
    [J]. ANALYTICAL CHEMISTRY, 1993, 65 (20) : 2933 - 2937
  • [8] SYNTHETIC CHEMORECEPTIVE MEMBRANES - SENSING BITTER OR ODOROUS SUBSTANCES ON A SYNTHETIC LIPID MULTIBILAYER FILM BY USING QUARTZ-CRYSTAL MICROBALANCES AND ELECTRIC RESPONSES
    OKAHATA, Y
    ENNA, G
    EBATO, H
    [J]. ANALYTICAL CHEMISTRY, 1990, 62 (14) : 1431 - 1438
  • [9] VERWENDUNG VON SCHWINGQUARZEN ZUR WAGUNG DUNNER SCHICHTEN UND ZUR MIKROWAGUNG
    SAUERBREY, G
    [J]. ZEITSCHRIFT FUR PHYSIK, 1959, 155 (02): : 206 - 222
  • [10] SI SH, 1994, ANAL LETT, V27, P2027