Mechanical performance comparison between RotaFlow and CentriMag centrifugal blood pumps in an adult ECLS model

被引:42
作者
Guan, Yulong [1 ]
Su, Xiaowei [1 ]
McCoach, Robert [1 ]
Kunselman, Allen [2 ]
El-Banayosy, Aly [3 ]
Uendar, Akif [1 ,4 ,5 ]
机构
[1] Penn State Childrens Hosp, Penn State Hershey Pediat Cardiovasc Res Ctr, Dept Pediat, Penn State Coll Med, Hershey, PA USA
[2] Penn State Hershey Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Hershey Coll Med, Hershey, PA USA
[3] Penn State Childrens Hosp, Inst Heart & Vasc, Penn State Milton S Hershey Med Ctr, Penn State Hershey Coll Med, Hershey, PA USA
[4] Penn State Childrens Hosp, Dept Surg, Penn State Milton S Hershey Med Ctr, Penn State Hershey Coll Med, Hershey, PA USA
[5] Penn State Childrens Hosp, Dept Bioengn, Penn State Milton S Hershey Med Ctr, Penn State Hershey Coll Med, Hershey, PA USA
来源
PERFUSION-UK | 2010年 / 25卷 / 02期
关键词
Extracorporeal life support; centrifugal blood pump; adults; EXTRACORPOREAL MEMBRANE-OXYGENATION; EXPERIENCE; SUPPORT; DEVICE;
D O I
10.1177/0267659110365366
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Centrifugal blood pumps have been widely adopted in conventional adult cardiopulmonary bypass and circulatory assist procedures. Different brands of centrifugal blood pumps incorporate distinct designs which affect pump performance. In this adult extracorporeal life support (ECLS) model, the performances of two brands of centrifugal blood pump (RotaFlow blood pump and CentriMag blood pump) were compared. Methods: The simulated adult ECLS circuit used in this study included a centrifugal blood pump, Quadrox D membrane oxygenator and Sorin adult ECLS tubing package. A Sorin Cardiovascular (R) VVR (R) 4000i venous reservoir (Sorin S.p.A., Milan, Italy) with a Hoffman clamp served as a pseudo-patient. The circuit was primed with 900ml heparinized human packed red blood cells and 300ml lactated Ringer's solution (total volume 1200 ml, corrected hematocrit 40%). Trials were conducted at normothermia (36 degrees C). Performance, including circuit pressure and flow rate, was measured for every setting analyzed. Results: The shut-off pressure of the RotaFlow was higher than the CentriMag at all measurement points given the same rotation speed (p < 0.0001). The shut-off pressure differential between the two centrifugal blood pumps was significant and increased given higher rotation speeds (p < 0.0001). The RotaFlow blood pump has higher maximal flow rate (9.08 +/- 0.01L/min) compared with the CentriMag blood pump (8.37 +/- 0.02L/min) (p < 0.0001). The blood flow rate differential between the two pumps when measured at the same revolutions per minute (RPM) ranged from 1.64L/min to 1.73L/min. Conclusions: The results obtained in this experiment demonstrate that the RotaFlow has a higher shut-off pressure (less retrograde flow) and maximal blood flow rate than the CentriMag blood pump. Findings support the conclusion that the RotaFlow disposable pump head has a better mechanical performance than the CentriMag. In addition, the RotaFlow disposable pump is 20-30 times less expensive than the CentriMag.
引用
收藏
页码:71 / 76
页数:6
相关论文
共 15 条
[1]   Bridge to life - Cardiac mechanical support [J].
Baughman, Kenneth L. ;
Jarcho, John A. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (09) :846-849
[2]   Pump-induced haemolysis: a comparison of short-term ventricular assist devices [J].
Bennett, M ;
Horton, S ;
Thuys, C ;
Augustin, S ;
Rosenberg, M ;
Brizard, C .
PERFUSION-UK, 2004, 19 (02) :107-111
[3]   Assessment of hydraulic performance and biocompatibility of a MagLev centrifugal pump system designed for pediatric cardiac or cardiopulmonary support [J].
Dasse, Kurt A. ;
Gellman, Barry ;
Kameneva, Marina V. ;
Woolley, Joshua R. ;
Johnson, Carl A. ;
Gempp, Thomas ;
Marks, John D. ;
Kent, Stella ;
Koert, Andrew ;
Richardson, J. Scott ;
Franklin, Steve ;
Snyder, Trevor A. ;
Wearden, Peter ;
Wagner, William R. ;
Gilbert, Richard J. ;
Borovetz, Harvey S. .
ASAIO JOURNAL, 2007, 53 (06) :771-777
[4]   Clinical performance with the levitronix centrimag short-term ventricular assist device [J].
De Robertis, F ;
Birks, EJ ;
Rogers, P ;
Dreyfus, G ;
Pepper, JR ;
Khaghani, A .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2006, 25 (02) :181-186
[5]   Extracorporeal membrane oxygenation with a poly-methylpentene oxygenator (Quadrox D). The experience of a single Italian centre in adult patients with refractory cardiogenic shock [J].
Formica, Francesco ;
Avalli, Leonello ;
Martino, Antonello ;
Maggioni, Elena ;
Muratore, Maria ;
Ferro, Orazio ;
Pesenti, Antonio ;
Paolini, Giovanni .
ASAIO JOURNAL, 2008, 54 (01) :89-94
[6]   Hydrodynamic performance and heat generation by centrifugal pumps [J].
Ganushchak, Y. ;
Lichtenbelt, W. van Marken ;
van der Nagel, T. ;
de Jong, D. S. .
PERFUSION-UK, 2006, 21 (06) :373-379
[7]   A magnetically levitated centrifugal blood pump with a simple-structured disposable pump head [J].
Hijikata, Wataru ;
Shinshi, Tadahiko ;
Asama, Junichi ;
Li, Lichuan ;
Hoshi, Hideo ;
Takatani, Setsuo ;
Shimokohbe, Akira .
ARTIFICIAL ORGANS, 2008, 32 (07) :531-540
[8]   Experience with the Jostra Rotaflow and QuadroxD oxygenator for ECMO [J].
Horton, S ;
Thuys, C ;
Bennett, M ;
Augustin, S ;
Rosenberg, M ;
Brizard, C .
PERFUSION-UK, 2004, 19 (01) :17-23
[9]   Early experience with the Levitronix Centrimag® device for extra-corporeal membrane oxygenation following lung transplantation [J].
Khan, Nouman U. ;
Al-Aloul, Mohamed ;
Shah, Rajesh ;
Yonan, Nizar .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2008, 34 (06) :1262-1264
[10]  
Lawson D Scott, 2005, Pediatr Crit Care Med, V6, P573, DOI 10.1097/01.PCC.0000163282.63992.13