Bactec 9240 blood culture system:: to preincubate at 35 °C or not?

被引:22
作者
Lemming, L
Holt, HM
Petersen, IS
Ostergaard, C
Bruun, B [1 ]
机构
[1] Hillerod Hosp, Dept Clin Microbiol, Hillerod, Denmark
[2] State Serum Inst, Dept Clin Microbiol, Copenhagen, Denmark
关键词
Bactec system; blood culture; preincubation of blood cultures;
D O I
10.1111/j.1469-0691.2004.00969.x
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Bactec Plus blood culture bottles were preincubated at 35degreesC or at room temperature before entry into the Bactec 9240 instrument to determine the influence of preincubation temperature and time. Of 463 positive blood culture sets, 956 bottles were positive, of which the instrument detected 92.1%. Of 76 positive bottles undetected by the instrument, 68 were preincubated at 35degreesC and eight at room temperature. The median entry delay and instrument detection times were 17.9 and 7.2 h for preincubated bottles, and 16.4 and 13.4 h for bottles held at room temperature. Short entry delay and inspection before entry into the instrument are necessary if preincubation at 35degreesC is used.
引用
收藏
页码:1089 / 1091
页数:3
相关论文
共 9 条
[1]  
*BECT DICK, 1998, BACTEC 9120 PLUS 924
[2]  
*BECT DICK, 1999, BACT FLUOR SER US MA
[3]  
*BECT DICK BIOSC, 1999, BACTEC TM 9000 SER D
[4]   Comparison of Bactec 9240 and Difco ESP blood culture systems for detection of organisms from vials whose entry was delayed [J].
Chapin, K ;
Lauderdale, TL .
JOURNAL OF CLINICAL MICROBIOLOGY, 1996, 34 (03) :543-549
[5]   CLINICAL IMPACT OF RAPID IN-VITRO SUSCEPTIBILITY TESTING AND BACTERIAL IDENTIFICATION [J].
DOERN, GV ;
VAUTOUR, R ;
GAUDET, M ;
LEVY, B .
JOURNAL OF CLINICAL MICROBIOLOGY, 1994, 32 (07) :1757-1762
[6]   COLORBACT, A VISUALLY READ BLOOD CULTURE SYSTEM USING DARKENING OF HEMOGLOBIN IN AEROBIC BLOOD CULTURES AS AN EARLY GROWTH INDICATOR, COMPARED WITH BACTEC-6A AND BACTEC-7A [J].
PRAG, J ;
JENSEN, J ;
LEBECH, K .
APMIS, 1991, 99 (12) :1089-1095
[7]   VALUE OF TERMINAL SUBCULTURES FOR BLOOD CULTURES MONITORED BY BACTEC-9240 [J].
SHIGEI, JT ;
SHIMABUKURO, JA ;
PEZZLO, MT ;
DELAMAZA, LM ;
PETERSON, EM .
JOURNAL OF CLINICAL MICROBIOLOGY, 1995, 33 (05) :1385-1388
[8]  
Weinstein MP, 1997, CLIN INFECT DIS, V24, P584, DOI 10.1093/clind/24.4.584
[9]   Controlled clinical laboratory comparison of two supplemented aerobic and anaerobic media used in automated blood culture systems to detect bloodstream infections [J].
Ziegler, R ;
Johnscher, I ;
Martus, P ;
Lenhardt, D ;
Just, HM .
JOURNAL OF CLINICAL MICROBIOLOGY, 1998, 36 (03) :657-661