An analytical dosimetry model as a step towards accounting for inhomogeneities and bounded geometries in 192Ir brachytherapy treatment planning

被引:33
作者
Anagnostopoulos, G
Baltas, D
Karaiskos, P
Pantelis, E
Papagiannis, P
Sakelliou, L [1 ]
机构
[1] Klinikum Offenbach, Strahlenklin, Dept Med Phys & Engn, D-63069 Offenbach, Germany
[2] Natl Tech Univ Athens, Inst Commun & Comp Sys, Athens 15773, Greece
[3] Univ Athens, Dept Phys, Nucl & Particle Phys Sect, Ilisia 15771, Greece
[4] Hygeia Hosp, Dept Phys Med, Athens 15123, Greece
关键词
D O I
10.1088/0031-9155/48/11/310
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
A simple analytical dose rate calculation model based on primary and scatter separation that treats 192 IF as a monoenergetic source by use of appropriate attenuation and mass energy absorption coefficients is documented for accurate dosimetry in water. This model is then generalized and tested for use in any homogeneous tissue material of radiobiological interest using scatter to primary ratios calculated in water with a material density scaling to account for the difference in the scattering properties of these materials and water. The potential of the analytical model for predicting the effect of the interference of an inhomogeneity is then evaluated by comparison with corresponding Monte Carlo calculations. It is found that regardless of the inhomogeneity dimensions and position relative to the source, the model is capable of increased accuracy (better than 2%) in calculating the primary dose rate at any point not only for low-Z tissue materials but also for high-Z shielding materials where a severe hardening of the primary photons occurs. Overall, for low-Z tissue inhomogeneities the proposed model succeeds in correcting dosimetry results towards the right direction compared to commercial treatment planning systems that currently ignore the effect of phantom dimensions and inhomogeneity interference. Regarding high-Z shielding materials the proposed model accurately predicts the dose reduction just beyond the inhomogeneity (for example it predicts a dose reduction of 47% just behind a tungsten alloy cylinder of 1 cm diameter and 2 mm thickness placed at 1.4 cm away from an Ir-192 source, in agreement with corresponding results in the literature) but does not account for the increasing contribution of the laterally scattered photons with increasing distance from the bounded inhomogeneity.
引用
收藏
页码:1625 / 1647
页数:23
相关论文
共 38 条
[1]
Monte Carlo dosimetry of a new 192Ir high dose rate brachytherapy source [J].
Angelopoulos, A ;
Baras, P ;
Sakelliou, L ;
Karaiskos, P ;
Sandilos, P .
MEDICAL PHYSICS, 2000, 27 (11) :2521-2527
[2]
ACCURATE MONTE-CARLO CALCULATIONS OF THE COMBINED ATTENUATION AND BUILDUP FACTORS, FOR ENERGIES (20-1500 KEV) AND DISTANCES (0-10 CM) RELEVANT IN BRACHYTHERAPY [J].
ANGELOPOULOS, A ;
PERRIS, A ;
SAKELLARIOU, K ;
SAKELLIOU, L ;
SARIGIANNIS, K ;
ZARRIS, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (06) :763-778
[3]
Comparison of calibration procedures for 192IR high-dose-rate brachytherapy sources [J].
Baltas, D ;
Geramani, K ;
Ioannidis, GT ;
Hierholz, K ;
Rogge, B ;
Kolotas, C ;
Müller-Sievers, K ;
Milickovic, N ;
Kober, B ;
Zamboglou, N .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (03) :653-661
[4]
Briesmeister J.F., 2000, LA13709M LOS AL NAT
[5]
Point kernels and superposition methods for scatter dose calculations in brachytherapy [J].
Carlsson, ÅK ;
Ahnesjö, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (02) :357-382
[6]
Dose rate constant and energy spectrum of interstitial brachytherapy sources [J].
Chen, Z ;
Nath, R .
MEDICAL PHYSICS, 2001, 28 (01) :86-96
[7]
Das RK, 1997, INT J RADIAT ONCOL, V38, P843, DOI 10.1016/S0360-3016(97)00067-9
[8]
Analytical approach to heterogeneity correction factor calculation for brachytherapy [J].
Daskalov, GM ;
Kirov, AS ;
Williamson, JF .
MEDICAL PHYSICS, 1998, 25 (05) :722-735
[9]
Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source [J].
Daskalov, GM ;
Loffler, E ;
Williamson, JF .
MEDICAL PHYSICS, 1998, 25 (11) :2200-2208
[10]
An analysis of MCNP cross-sections and tally methods for low-energy photon emitters [J].
DeMarco, JJ ;
Wallace, RE ;
Boedeker, K .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (08) :1321-1332