Artificial composite bone as a model of human trabecular bone:: The implant-bone interface

被引:136
作者
Grant, J. A.
Bishop, N. E.
Goetzen, N.
Sprecher, C.
Honl, M.
Morlock, M. M.
机构
[1] TUHH Hamburg Univ Technol, Biomech Sect, D-21073 Hamburg, Germany
[2] AO Res Inst, Davos, Switzerland
[3] Rush Univ, Ctr Med, Chicago, IL 60612 USA
关键词
friction; bone-implant interface; artificial bone; pre-clinical testing; press fit;
D O I
10.1016/j.jbiomech.2006.04.007
中图分类号
Q6 [生物物理学];
学科分类号
071011 [生物物理学];
摘要
The use of artificial bones in implant testing has become popular due to their low variability and ready availability. However, friction coefficients, which are critical to load transfer in uncemented implants, have rarely been compared between human and artificial bone, particularly for wet and dry conditions. In this study, the static and dynamic friction coefficients for four commercially used titanium surfaces (polished, Al2O3 blasted, plasma sprayed, beaded) acting on the trabecular component of artificial bones (Sawbones (R)) were compared to those for human trabecular bone. Artificial bones were tested in dry and wet conditions and normal interface stress was varied (0.25, 0.5, 1.0 MPa). Friction coefficients were mostly lower for artificial bones than real bone. In particular, static friction coefficients for the dry polished surface were 20 % of those for real bone and 42-61 % for the dry beaded surface, with statistical significance (alpha<0.05). Less marked differences were observed for dynamic friction coefficients. Significant but non-systematic effects of normal stress or wet/dry condition on friction coefficients were observed within each surface type. These results indicate that the use of artificial bone models for pre-clinical implant testing that rely on interface load transfer with trabecular bone for mechanical integrity can be particularly sensitive to surface finish and lubrication conditions. (c) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1158 / 1164
页数:7
相关论文
共 8 条
[1]
INITIAL STABILITY OF UNCEMENTED HIP STEMS - AN IN-VITRO PROTOCOL TO MEASURE TORSIONAL INTERFACE MOTION [J].
HARMAN, MK ;
TONI, A ;
CRISTOFOLINI, L ;
VICECONTI, M .
MEDICAL ENGINEERING & PHYSICS, 1995, 17 (03) :163-171
[2]
COMPARISON OF THE STABILITY OF PRESS-FIT HIP-PROSTHESIS FEMORAL STEMS USING A SYNTHETIC MODEL FEMUR [J].
MCKELLOP, H ;
EBRAMZADEH, E ;
NIEDERER, PG ;
SARMIENTO, A .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1991, 9 (02) :297-305
[3]
Methods for quantitative analysis of the primary stability in uncemented hip prostheses [J].
Monti, L ;
Cristofolini, L ;
Viceconti, M .
ARTIFICIAL ORGANS, 1999, 23 (09) :851-859
[4]
On the secondary stability of coated cementless hip replacement: parameters that affected interface strength [J].
Orlik, J ;
Zhurov, A ;
Middleton, J .
MEDICAL ENGINEERING & PHYSICS, 2003, 25 (10) :825-831
[5]
Otani T, 1993, J Arthroplasty, V8, P67, DOI 10.1016/S0883-5403(06)80110-5
[6]
FRICTIONAL INTERFACE MICROMOTIONS AND ANISOTROPIC STRESS-DISTRIBUTION IN A FEMORAL TOTAL HIP COMPONENT [J].
RUBIN, PJ ;
RAKOTOMANANA, RL ;
LEYVRAZ, PF ;
ZYSSET, PK ;
CURNIER, A ;
HEEGAARD, JH .
JOURNAL OF BIOMECHANICS, 1993, 26 (06) :725-739
[7]
EXPERIMENTAL-DETERMINATION OF FRICTION CHARACTERISTICS AT THE TRABECULAR BONE POROUS-COATED METAL INTERFACE IN CEMENTLESS IMPLANTS [J].
SHIRAZIADL, A ;
DAMMAK, M ;
PAIEMENT, G .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (02) :167-175
[8]
Pre-clinical validation of a new partially cemented femoral prosthesis by synergetic use of numerical and experimental methods [J].
Viceconti, M ;
Cristofolini, L ;
Baleani, M ;
Toni, A .
JOURNAL OF BIOMECHANICS, 2001, 34 (06) :723-731