Noninvasive motion ventilation (NIMV): a novel approach to ventilatory support

被引:31
作者
Adams, JA
Mangino, MJ
Bassuk, J
Inman, DM
Sackner, MA
机构
[1] Mt Sinai Med Ctr, Div Neonatol, Miami Beach, FL 33140 USA
[2] Mt Sinai Med Ctr, Dept Res, Miami Beach, FL 33140 USA
[3] Noninvas Monitoring Syst, Miami Beach, FL 33139 USA
[4] Mt Sinai Med Ctr, Div Pulm Dis, Miami Beach, FL 33140 USA
关键词
inertial forces; low-pressure respiratory airflow; pulmonary mechanics; low-tidal-volume ventilation;
D O I
10.1152/jappl.2000.89.6.2438
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
A motion platform was developed that oscillates an animal in a foot-to-head direction (z-plane). The platform varies the frequency and intensity of acceleration, imparting periodic sinusoidal inertial forces (pG(z)) to the body. The aim of the study was to characterize ventilation produced by the noninvasive motion ventilator (NIMV) in animals with healthy and diseased lungs. Incremental increases in pG(z) (acceleration) with the frequency held constant (f = 4 Hz) produced almost linear increases in minute ventilation (VE). Frequencies of 2-4 Hz produced the greatest V(over dot)(E) and tidal volume (VT) for any given acceleration between +/-0.2 and +/-0.8 G. Increasing the force due to acceleration produced proportional increases in both transpulmonary and transdiaphragmatic pressures. Increasing transpulmonary pressure by increasing pG(z) produced linear increases in V-T, similar to spontaneous breathing. NIMV reversed deliberately induced hypoventilation and normalized the changes in arterial blood gases induced by meconium aspiration. In conclusion, a novel motion platform is described that imparts periodic sinusoidal acceleration forces at moderate frequencies (4 Hz) to the whole body in the z-plane. These forces, when properly adjusted, are capable of highly effective ventilation of normal and diseased lungs. Such noninvasive ventilation is accomplished at airway pressures equivalent to atmospheric or continuous positive airway pressure, with acceleration forces less than +/-1 G(z).
引用
收藏
页码:2438 / 2446
页数:9
相关论文
共 21 条
[1]   Hemodynamic effects of periodic Gz acceleration in meconium aspiration in pigs [J].
Adams, JA ;
Mangin, MJ ;
Bassuk, J ;
Sackner, MA .
JOURNAL OF APPLIED PHYSIOLOGY, 2000, 89 (06) :2447-2452
[2]   THORACOABDOMINAL ASYNCHRONY IN INFANTS WITH AIR-FLOW OBSTRUCTION [J].
ALLEN, JL ;
WOLFSON, MR ;
MCDOWELL, K ;
SHAFFER, TH .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1990, 141 (02) :337-342
[3]  
BAUDINETTE RV, 1987, J EXP BIOL, V129, P251
[4]  
BAYDUR A, 1982, AM REV RESPIR DIS, V126, P788
[5]   REMOVAL OF BRONCHIAL-SECRETIONS BY 2-PHASE GAS-LIQUID TRANSPORT [J].
BENJAMIN, RG ;
CHAPMAN, GA ;
KIM, CS ;
SACKNER, MA .
CHEST, 1989, 95 (03) :658-663
[6]  
COERMANN ROLF R., 1960, AEROSPACE MED, V31, P443
[7]   MEASUREMENT OF TIDAL VOLUME USING A PNEUMOTACHOMETER DURING HIGH-FREQUENCY OSCILLATION [J].
COURTNEY, SE ;
WEBER, KR ;
SPOHN, WA ;
MALIN, SW ;
BENDER, CV ;
GOTSHALL, RW .
CRITICAL CARE MEDICINE, 1990, 18 (06) :651-653
[8]   Chest vibration redistributes intra-airway CO2 during tracheal insufflation in ventilatory failure [J].
Eckmann, DM ;
Gavriely, N .
CRITICAL CARE MEDICINE, 1996, 24 (03) :451-457
[9]   COMPARATIVE-STUDY OF INTRA-AIRWAY GAS-TRANSPORT BY ALTERNATIVE MODES OF VENTILATION [J].
GAVRIELY, N ;
GAVER, DP ;
SOLWAY, J ;
GROTBERG, JB .
JOURNAL OF APPLIED PHYSIOLOGY, 1995, 79 (05) :1512-1518
[10]   DIAPHRAGMATIC WORK OF BREATHING IN PREMATURE HUMAN INFANTS [J].
GUSLITS, BG ;
GASTON, SE ;
BRYAN, MH ;
ENGLAND, SJ ;
BRYAN, AC .
JOURNAL OF APPLIED PHYSIOLOGY, 1987, 62 (04) :1410-1415