A simple method for investigating the effects of non-uniformity of radiofrequency transmission and radiofrequency reception in MRI

被引:57
作者
Barker, GJ
Simmons, A
Arridge, SR
Tofts, PS
机构
[1] Inst Neurol, NMR Res Grp, London WC1N 3BG, England
[2] UCL, Dept Med Phys, London, England
[3] UCL, Dept Comp Sci, London, England
关键词
D O I
10.1259/bjr.71.841.9534700
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Inhomogeneity of the transmitted or received B(1) field leads to intensity variations in MR images and spatial dependence in apparent concentration in MR spectra. We describe a simple method for investigating such variations. The transmitted B(1) field can be measured both in vivo and in vitro which allows investigation of sample dependent effects that can not be measured on phantoms. For homogeneous regions the method also allows the received B(1) field to be measured both in vivo and in vitro. Our method uses only a standard spin echo pulse sequence and simple region of interest analysis and should be implementable on any commercial scanner. The method is demonstrated using a variety of transmission and reception radiofrequency coils both in vivo and in vitro.
引用
收藏
页码:59 / 67
页数:9
相关论文
共 17 条
  • [11] Plummer D., 1992, RIV NEURORADIOL, V5, P489, DOI DOI 10.1177/197140099200500413
  • [12] THE NMR PHASED-ARRAY
    ROEMER, PB
    EDELSTEIN, WA
    HAYES, CE
    SOUZA, SP
    MUELLER, OM
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1990, 16 (02) : 192 - 225
  • [13] SOURCES OF INTENSITY NONUNIFORMITY IN SPIN-ECHO IMAGES AT 1.5-T
    SIMMONS, A
    TOFTS, PS
    BARKER, GJ
    ARRIDGE, SR
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (01) : 121 - 128
  • [14] STANDING WAVES IN UNIFORM WATER PHANTOMS
    TOFTS, PS
    [J]. JOURNAL OF MAGNETIC RESONANCE SERIES B, 1994, 104 (02): : 143 - 147
  • [15] CORRECTION OF NONUNIFORMITY IN IMAGES OF THE SPINE AND OPTIC-NERVE FROM FIXED RECEIVE-ONLY SURFACE COILS AT 1.5-T
    TOFTS, PS
    BARKER, GJ
    SIMMONS, A
    MACMANUS, DG
    THORPE, J
    GASS, A
    MILLER, DH
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1994, 18 (06) : 997 - 1003
  • [16] PREPARATION OF AGAROSE GELS AS REFERENCE SUBSTANCES FOR NMR RELAXATION-TIME MEASUREMENT
    WALKER, P
    LERSKI, RA
    MATHURDEVRE, R
    BINET, J
    YANE, F
    [J]. MAGNETIC RESONANCE IMAGING, 1988, 6 (02) : 215 - 222
  • [17] CORRECTION OF INTENSITY NONUNIFORMITY IN MR IMAGES OF ANY ORIENTATION
    WICKS, DAG
    BARKER, GJ
    TOFTS, PS
    [J]. MAGNETIC RESONANCE IMAGING, 1993, 11 (02) : 183 - 196