Determination of arsenic compounds by high-performance liquid chromatography-ultrasonic nebulizer-high power nitrogen-microwave-induced plasma mass spectrometry: An accepted coupling

被引:27
作者
Chatterjee, A [1 ]
Shibata, Y [1 ]
Yoshinaga, J [1 ]
Morita, M [1 ]
机构
[1] Natl Inst Environm Studies, Environm Chemodynam Sect, Environm Chem Div, Tsukuba, Ibaraki 3050053, Japan
关键词
D O I
10.1021/ac000311s
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
To establish a sensitive, accurate, and precise determination of arsenic compounds, a high power nitrogen microwave-induced plasma (1.3 kW) mass spectrometer (N-2-MIP-MS) has been successfully coupled with an ultrasonic nebulizer (HPLC-USN) that is attached to a high-performance liquid chromatograph. It is examined as an element-specific detector for its applicability to the optimization and determination of seven arsenic compounds [arsenic acid, methylarsonic acid(MA), dimethylarsinic acid (DMA), arsenobetaine (AB), arsenocholine (AC), trimethylarsine oxide (TMAO), and tetramethylarsonium ion (TMI)]. This HPLC-USN-MIP-MS coupling is an encouraging combination as an alternative method for mass spectroscopy for elemental speciation analysis. Interchanging of the MIP-MS fabricated nebulizer (concentric) with an ultrasonic nebulizer, increases 3-6 times the ion signals for the anionic and 6-12 times those for the cationic arsenic compounds as compared to traditional methods. The HPLC-USN-MIP-MS combination used is excellent, amplifying the ion signals about 1.5-2 times for cationic and 1.3-3.8 times for the anionic arsenic compounds as compared to the HPLC-ICPMS coupling, The detection limits for As(V), MA, DMA, AB, TMAO, AC, and TMI (in Milli-Q-water) obtained with the optimized HPLC-USN-N-2-MIP-MS system are 0.46, 0.36, 0.73, 0.21, 3.64, 0.39, and 0.32 mu g L-1, respectively, about 13-50 times lower than the HPLC-MIP-MS and about 3-11 times lower than the HPLC-ICPMS. The detection limits of As(V), MA, DMA, AB, TMAO, AC, and TMI, which spike in the urine, are deteriorated by 1.7-4.2 times compared with the detection limits of the seven different As compounds, which are prepared in the Milli-Q-water, The repeatability (RSD for three successive analyses) and reproducibility (RSD for three successive analyses performed on three different days), considering peak area and peak height, achieved for seven different arsenic compounds are 0.5-7 and 0.7-8%, comparable with the HPLC-ICPMS (0.3-8.5%; 4-12%) and HPLC-MIP-MS (0.4-9%; 5-12%) systems. The combined HPLC-USN-N-2-MIP-MS has been adequately applied to the determination of AB in NIES Candidate Human Urine CRM. The results agree with the HPLC-ICPMS values. Chloride interference as (ArCl+)-Ar-40-Cl-35 is not found in the urine and with the high chloride matrix (10 000 mg L-1).
引用
收藏
页码:4402 / 4412
页数:11
相关论文
共 51 条
[1]  
ANDERSON J, 1992, ATOM SPECTROSC, V13, P99
[2]   APPLICATIONS OF ULTRASONIC NEBULIZATION IN THE ANALYSIS OF PETROLEUM AND PETROCHEMICALS BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
BOTTO, RI .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (01) :51-57
[3]   USE OF AN ULTRASONIC NEBULIZER WITH MEMBRANE DESOLVATION FOR ANALYSIS OF VOLATILE SOLVENTS BY INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
BOTTO, RI ;
ZHU, JJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (09) :905-912
[4]   DETERMINATION OF ARSENIC IN SAMPLES WITH HIGH CHLORIDE CONTENT BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BRANCH, S ;
EBDON, L ;
FORD, M ;
FOULKES, M ;
ONEILL, P .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (02) :151-154
[5]   PERFORMANCE-CHARACTERISTICS OF AN ULTRASONIC NEBULIZER COUPLED TO A 40.68 MHZ INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECTROMETER [J].
BRENNER, IB ;
BREMIER, P ;
LEMARCHAND, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (06) :819-824
[6]   Matrix effects in inductively coupled plasma atomic emission spectrometry using an ultrasonic nebulizer [J].
Budic, B .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (09) :869-874
[7]   EVALUATION OF AN ULTRASONIC NEBULIZER FOR SAMPLE INTRODUCTION IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
CASTILLANO, TM ;
VELA, NP ;
CARUSO, JA ;
STORY, WC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (06) :807-811
[8]   Application of a nitrogen microwave-induced plasma mass spectrometer as an element-specific detector for arsenic speciation analysis [J].
Chatterjee, A ;
Shibata, Y ;
Yoshinaga, J ;
Morita, M .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (12) :1853-1859
[9]   Behaviour of anionic arsenic compounds in microwave system with nitric acid and hydrogen peroxide - preliminary laboratory study [J].
Chatterjee, A .
SCIENCE OF THE TOTAL ENVIRONMENT, 1999, 228 (01) :25-34
[10]   ARSENIC IN GROUND-WATER IN 6 DISTRICTS OF WEST-BENGAL, INDIA - THE BIGGEST ARSENIC CALAMITY IN THE WORLD .1. ARSENIC SPECIES IN DRINKING-WATER AND URINE OF THE AFFECTED PEOPLE [J].
CHATTERJEE, A ;
DAS, D ;
MANDAL, BK ;
CHOWDHURY, TR ;
SAMANTA, G ;
CHAKRABORTI, D .
ANALYST, 1995, 120 (03) :643-650