Calculating output factors for photon beam radiotherapy using a convolution/superposition method based on a dual source photon beam model

被引:48
作者
Liu, HH [1 ]
Mackie, TR
McCullough, EC
机构
[1] Mayo Clin & Mayo Fdn, Div Radiat Oncol, Rochester, MN 55905 USA
[2] Univ Wisconsin, Dept Med Phys, Madison, WI 53706 USA
关键词
convolution/superposition; Monte Carlo simulation; extra-focal radiation; photon output factor; head scatter factor; phantom scatter factor; jaw exchange effect; asymmetric fields; multileaf collimator;
D O I
10.1118/1.598111
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A realistic photon beam model based on Monte Carlo simulation of clinical linear accelerators was implemented in a convolution/superposition dose calculation algorithm. A primary and an extrafocal sources were used in this beam model to represent the direct photons from the target and the scattered photons from other head structures, respectively. The effect of the finite size of the extra-focal source was modeled by a convolution of the source fluence distribution with the collimator aperture function. Relative photon output in air (S-c) and in phantom (S-cp) were computed using the convolution method with this new photon beam model. Our results showed that in a 10 MV photon beam, the S-c, S-p (phantom scatter factor), and S-cp factors increased by 11%, 10%, and 22%, respectively, as the field size changed from 3 x 3 cm(2) to 40 x 40 cm(2). The variation of the Sc factor was contributed mostly by an increase of the extra-focal radiation with field size. The radiation backscattered into the monitor chamber inside the accelerator head affected the S-c by about 2% in the same field range. The output factors in elongated fields, asymmetric fields, and blocked fields were also investigated in this study. Our results showed that if the effect of the backscattered radiation was taken into account, output factors in these treatment fields can be predicted accurately by our convolution algorithm using the dual source photon beam model. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:1975 / 1985
页数:11
相关论文
共 32 条
[1]   A PENCIL BEAM MODEL FOR PHOTON DOSE CALCULATION [J].
AHNESJO, A ;
SAXNER, M ;
TREPP, A .
MEDICAL PHYSICS, 1992, 19 (02) :263-273
[2]   APPLICATION OF THE CONVOLUTION METHOD FOR CALCULATION OF OUTPUT FACTORS FOR THERAPY PHOTON BEAMS [J].
AHNESJO, A ;
KNOOS, T ;
MONTELIUS, A .
MEDICAL PHYSICS, 1992, 19 (02) :295-301
[3]   COLLAPSED CONE CONVOLUTION OF RADIANT ENERGY FOR PHOTON DOSE CALCULATION IN HETEROGENEOUS MEDIA [J].
AHNESJO, A .
MEDICAL PHYSICS, 1989, 16 (04) :577-592
[4]   MONTE-CARLO STUDY OF ACCELERATOR HEAD SCATTER [J].
CHANEY, EL ;
CULLIP, TJ ;
GABRIEL, TA .
MEDICAL PHYSICS, 1994, 21 (09) :1383-1390
[5]  
DEAN EM, 1993, BRIT J RADIOL, V66, P517
[6]   ON THE FIELD-SIZE DEPENDENCE OF RELATIVE OUTPUT FROM A LINEAR-ACCELERATOR [J].
DUNSCOMBE, PB ;
NIEMINEN, JM .
MEDICAL PHYSICS, 1992, 19 (06) :1441-1444
[7]   BACKSCATTER INTO THE BEAM MONITOR CHAMBER - IMPLICATIONS FOR DOSIMETRY OF ASYMMETRIC COLLIMATORS [J].
DUZENLI, C ;
MCCLEAN, B ;
FIELD, C .
MEDICAL PHYSICS, 1993, 20 (02) :363-367
[8]   SCATTER FACTOR CORRECTIONS FOR ELONGATED FIELDS [J].
HIGGINS, PD ;
SOHN, WH ;
SIBATA, CH ;
MCCARTHY, WA .
MEDICAL PHYSICS, 1989, 16 (05) :800-802
[9]   THE EFFECT OF COLLIMATOR BACKSCATTER RADIATION ON PHOTON OUTPUT OF LINEAR ACCELERATORS [J].
HUANG, PH ;
CHU, J ;
BJARNGARD, BE .
MEDICAL PHYSICS, 1987, 14 (02) :268-269
[10]   X-RAY SOURCES OF MEDICAL LINEAR ACCELERATORS - FOCAL AND EXTRA-FOCAL RADIATION [J].
JAFFRAY, DA ;
BATTISTA, JJ ;
FENSTER, A ;
MUNRO, P .
MEDICAL PHYSICS, 1993, 20 (05) :1417-1427