In vivo estimates of the position of advanced bionics electrode arrays in the human cochlea

被引:203
作者
Skinner, Margaret W.
Holden, Timothy A.
Whiting, Bruce R.
Voie, Arne H.
Brunsden, Barry
Neely, J. Gail
Saxon, Eugene A.
Hullar, Timothy E.
Finley, Charles C.
机构
[1] Washington Univ, Sch Med, Dept Otolaryngol Head & Neck Surg, St Louis, MO 63110 USA
[2] Spencer Technol, Seattle, WA USA
[3] Univ N Carolina, Sch Med, Dept Otolaryngol Head & Neck Surg, Chapel Hill, NC USA
关键词
cochlear implant; electrode array; image volume registration; spiral computed tomography; three-dimensional reconstruction; word recognition;
D O I
10.1177/000348940711600401
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Objectives: A new technique for determining the position of each electrode in the cochlea is described and applied to spiral computed tomography data from 15 patients implanted with Advanced Bionics HiFocus I, Ij, or Helix arrays. Methods: ANALYZE imaging software was used to register 3-dimensional image volumes from patients' preoperative and postoperative scans and from a single body donor whose unimplanted ears were scanned clinically, with micro computed tomography and with orthogonal-plane fluorescence optical sectioning (OPFOS) microscopy. By use of this registration, we compared the atlas of OPFOS images of soft tissue within the body donor's cochlea with the bone and fluid/tissue boundary available in patient scan data to choose the midmodiolar axis position and judge the electrode position in the scala tympani or scala vestibuli, including the distance to the medial and lateral scalar walls. The angular rotation 0 degrees start point is a line joining the midmodiolar axis and the middle of the cochlear canal entry from the vestibule. Results: The group mean array insertion depth was 477 degrees (range, 286 degrees to 655 degrees). The word scores were negatively correlated (r = -0.59; p = .028) with the number of electrodes in the scala vestibuli. Conclusions: Although the individual variability in all measures was large, repeated patterns of suboptimal electrode placement were observed across subjects, underscoring the applicability of this technique.
引用
收藏
页码:2 / 24
页数:23
相关论文
共 77 条
[1]  
Adunka O, 2005, ARCH OTOLARYNGOL, V131, P488, DOI 10.1001/archotol.131.6.488
[2]  
Advanced Bionics Corporation, 2005, HIRES 90K SURG MAN H
[3]   Quality control after cochlear implant surgery by means of rotational tomography [J].
Aschendorff, A ;
Kubalek, T ;
Turowski, B ;
Zanella, T ;
Hochmuth, T ;
Schumacher, T ;
Klenzner, T ;
Laszig, R .
OTOLOGY & NEUROTOLOGY, 2005, 26 (01) :34-37
[4]   Imaging procedures in cochlear implant patients - Evaluation of different radiological techniques [J].
Aschendorff, A ;
Kubalek, R ;
Hochmuth, A ;
Bink, A ;
Kurtz, C ;
Lohnstein, P ;
Klenzner, T ;
Laszig, R .
ACTA OTO-LARYNGOLOGICA, 2004, 124 :46-49
[5]   Auditory cortical images of cochlear-implant stimuli: Dependence on electrode configuration [J].
Bierer, JA ;
Middlebrooks, JC .
JOURNAL OF NEUROPHYSIOLOGY, 2002, 87 (01) :478-492
[6]   Acoustic to electric pitch comparisons in cochlear implant subjects with residual hearing [J].
Boëx, Colette ;
Baud, Lionel ;
Cosendai, Grégoire ;
Sigrist, Alain ;
Kós, Maria-Izabel ;
Pelizzone, Marco .
JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2006, 7 (02) :110-124
[7]  
BREDBERG G, 1968, ACTA OTOLARYNGOL S, V236
[8]   A novel binning method for improved accuracy and speed of volume image coregistration using normalized mutual information [J].
Camp, J ;
Robb, R .
MEDICAL IMAGING 1999: IMAGE PROCESSING, PTS 1 AND 2, 1999, 3661 :24-31
[9]  
Chen JM, 1999, AM J OTOL, V20, P192
[10]  