Causes of failure of the Kinemax Plus total knee replacement in the first five years

被引:6
作者
Davis, C. R. [1 ]
Davies, A. P. [1 ]
Newman, J. H. [1 ]
机构
[1] Southmead Gen Hosp, Avon Orthopaed Ctr, Bristol BS10 5NB, Avon, England
关键词
total knee replacement; Kinemax Plus; revision; cause of failure; polyethylene;
D O I
10.1016/j.knee.2007.01.002
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
In order to investigate whether any single batch of polyethylene inserts was responsible for premature failure of total knee replacements (TKR), we performed a retrospective database review of 1133 consecutive primary Kinemax Plus TKR performed by 2 surgical teams between January 1992 and January 2001 at our institution. Cases re-operated in any way within 5 years of the index arthroplasty were identified, case notes studied and radiographs reviewed. The causes contributing to the premature failure of the implants were identified where possible and the polyethylene lot numbers were recorded. Nineteen cases were known to have been revised, 7 cases for deep infection of the prosthesis and 12 for aseptic failure. In 10 cases, an identifiable technical error had contributed to the early failure of the implant. In 4 of these cases, there was inadequate tissue balancing in patients with a significant pre-operative valgus deformity. In 2 cases, the tibial base plate was inserted in a varus mal-alignment. There was one extended tibial component, one internally rotated femoral component, one oversized femoral component and one femoral component with inadequate fixation. Two further cases failed aseptically, one with a fracture of the medial tibial plateau and one idiopathic. The lot numbers of the polyethylene inserts were all different. This study confirms that the premature failure rate of the Kinemax Plus TKR is low. Although a "bad batch" of polyethylene has been suggested as a cause for early failure in TKR, in this series, surgical error was the predominant cause. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:158 / 161
页数:4
相关论文
共 13 条
[1]
Medium-term results of the Scorpio total knee replacement [J].
Abbas, Dawar ;
Gunn, R. S. .
KNEE, 2006, 13 (04) :307-311
[2]
The Kinemax total knee arthroplasty - Nine years' experience [J].
Back, DL ;
Cannon, SR ;
Hilton, A ;
Bankes, MJK ;
Briggs, TWR .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (03) :359-363
[3]
The effect of component malalignment on the clinical and radiological outcome of the Kinemax total knee replacement [J].
Bankes, MJK ;
Back, DL ;
Cannon, SR ;
Briggs, TWR .
KNEE, 2003, 10 (01) :55-60
[4]
Five-year results of the Sigma total knee arthroplasty [J].
Clayton, Robert A. E. ;
Amin, Anish K. ;
Gaston, Mark S. ;
Brenkel, Ivan J. .
KNEE, 2006, 13 (05) :359-364
[5]
Infecting organisms in primary total joint replacement [J].
Davies, A. P. .
JOURNAL OF HOSPITAL INFECTION, 1998, 38 (01) :70-70
[6]
Total knee arthroplasty for severe valgus deformity - Five to fourteen-year follow-up [J].
Elkus, M ;
Ranawat, CS ;
Rasquinha, VJ ;
Babhulkar, S ;
Rossi, R ;
Ranawat, AS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (12) :2671-2676
[7]
Fehring TK, 2001, CLIN ORTHOP RELAT R, P315
[8]
Revision rates after knee replacement in the United States [J].
Heck, DA ;
Melfi, CA ;
Mamlin, LA ;
Katz, BP ;
Arthur, DS ;
Dittus, RS ;
Freund, DA .
MEDICAL CARE, 1998, 36 (05) :661-669
[9]
SEVERE VARUS AND VALGUS DEFORMITIES TREATED BY TOTAL KNEE ARTHROPLASTY [J].
KARACHALIOS, T ;
SARANGI, PP ;
NEWMAN, JH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (06) :938-942
[10]
Total knee replacement using the Genesis II prosthesis: a 5-year follow up study of the first 100 consecutive cases [J].
Laskin, RS ;
Davis, J .
KNEE, 2005, 12 (03) :163-167