Current calibration, treatment, and treatment planning techniques among institutions participating in the Children's Oncology Group

被引:16
作者
Urie, M
FitzGerald, TJ
Followill, D
Laurie, F
Marcus, R
Michalski, J
机构
[1] Qual Assurance Review Ctr, Providence, RI 02908 USA
[2] Univ Massachusetts, Sch Med, Worcester, MA USA
[3] Univ Texas, MD Anderson Canc Ctr, Houston, TX 77030 USA
[4] Emory Univ, Sch Med, Atlanta, GA USA
[5] Washington Univ, Mallinckrodt Inst Radiol, St Louis, MO USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2003年 / 55卷 / 01期
关键词
D O I
10.1016/S0360-3016(02)03827-0
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To report current technology implementation, radiation therapy physics and treatment planning practices, and results of treatment planning exercises among 261 institutions belonging to the Children's Oncology Group (COG). Methods and Materials: The Radiation Therapy Committee of the newly formed COG mandated that each institution demonstrate basic physics and treatment planning abilities by satisfactorily completing a questionnaire and four treatment planning exercises designed by the Quality Assurance Review Center. The planning cases are (1) a maxillary sinus target volume (for two-dimensional planning), (2) a Hodgkin's disease mantle field (for irregular-field and off-axis dose calculations), (3) a central axis blocked case, and (4) a craniospinal irradiation case. The questionnaire and treatment plans were submitted (as of 1/30/02) by 243 institutions and completed satisfactorily by 233. Data from this questionnaire and analyses of the treatment plans with monitor unit calculations are presented. Results: Of the 243 clinics responding, 54% use multileaf collimators routinely, 94% use asymmetric jaws routinely, and 13% use dynamic wedges. Nearly all institutions calibrate their linear accelerators following American Association of Physicists in Medicine protocols, currently 16% with TG-51 and 81% with TG-21 protocol. Treatment planning systems are relied on very heavily for all calculations, including monitor units. Techniques and results of each of the treatment planning exercises are presented. Conclusions: Together, these data provide a unique compilation of current (2001) radiation therapy practices in institutions treating pediatric patients. Overall, the COG facilities have the equipment and the personnel to perform high-quality radiation therapy. With ongoing quality assurance review, radiation therapy compliance with COG protocols should be high. (C) 2003 Elsevier Science Inc.
引用
收藏
页码:245 / 260
页数:16
相关论文
共 21 条
[1]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[2]  
[Anonymous], 1993, 50 ICRU
[3]  
[Anonymous], 1999, 62 ICRU
[4]   Clinical impact of dosimetry quality assurance programmes assessed by radiobiological modelling of data from the thermoluminescent dosimetry study of the European Organization for Research and Treatment of Cancer [J].
Bentzen, SM ;
Bernier, J ;
Davis, JB ;
Horiot, JC ;
Garavaglia, G ;
Chavaudra, J ;
Johansson, KA ;
Bolla, M .
EUROPEAN JOURNAL OF CANCER, 2000, 36 (05) :615-620
[5]   QUALITY ASSURANCE OF THE EORTC RADIOTHERAPY TRIAL-22863 FOR PROSTATIC-CANCER - THE DUMMY RUN [J].
DUSSERRE, A ;
GARAVAGLIA, G ;
GIRAUD, JY ;
BOLLA, M .
RADIOTHERAPY AND ONCOLOGY, 1995, 36 (03) :229-234
[6]  
FOLLOWILL D, 2002, COMMUNICATION APR
[7]  
Gagnon W F, 1978, Med Phys, V5, P556, DOI 10.1118/1.594450
[8]   QUALITY ASSURANCE CONTROL IN THE EORTC COOPERATIVE GROUP OF RADIOTHERAPY .3. INTERCOMPARISON IN AN ANATOMICAL PHANTOM [J].
JOHANSSON, KA ;
HORIOT, JC ;
VANDERSCHUEREN, E .
RADIOTHERAPY AND ONCOLOGY, 1987, 9 (04) :289-298
[9]   QUALITY ASSURANCE CONTROL IN THE EORTC COOPERATIVE GROUP OF RADIOTHERAPY .2. DOSIMETRIC INTERCOMPARISON [J].
JOHANSSON, KA ;
HORIOT, JC ;
VANDAM, J ;
LEPINOY, D ;
SENTENAC, I ;
SERNBO, G .
RADIOTHERAPY AND ONCOLOGY, 1986, 7 (03) :269-279
[10]   ELECTRON-BEAM CENTRAL AXIS DEPTH DOSE MEASUREMENTS [J].
KIRBY, TH ;
GASTORF, RJ ;
HANSON, WF ;
BERKLEY, LW ;
GAGNON, WF ;
HAZLE, JD ;
SHALEK, RJ .
MEDICAL PHYSICS, 1985, 12 (03) :357-361