A Mediated Glucose Biosensor Incorporated with Reverse Iontophoresis Function for Noninvasive Glucose Monitoring

被引:16
作者
Ching, Congo Tak-Shing [1 ,2 ]
Sun, Tai-Ping [1 ,2 ]
Huang, Su-Hua [3 ]
Shieh, Hsiu-Li [2 ]
Chen, Chung-Yuan [2 ]
机构
[1] Natl Chi Nan Univ, Grad Inst Biomed & Biomed Technol, Puli 54561, Nantou County, Taiwan
[2] Natl Chi Nan Univ, Dept Elect Engn, Puli 54561, Nantou County, Taiwan
[3] Asia Univ, Dept Biotechnol, Wufeng 41354, Taichung County, Taiwan
关键词
Amperometric; Ferrocene; Glucose; Biosensors; Reverse Iontophoresis; AMPEROMETRIC ENZYME ELECTRODE; GLUCOWATCH BIOGRAPHER; ASCORBIC-ACID; IN-VIVO; FERROCENE; OXIDASE; DESIGN; FLOW; TETRATHIAFULVALENE; COMPOSITE;
D O I
10.1007/s10439-010-9918-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The objective of this study was to develop a ferrocene mediated glucose biosensor for reverse iontophoresis. An amperometric ferrocene mediated glucose biosensor based on a three electrodes planar configuration was constructed using screen printing technique. Different combinations of glucose oxidase and ferrocene loading were drop coated onto the surface of the amperometric transducer. The amperometric transducer was characterized electrochemically using cyclic voltammetry and its electrochemical characteristics (Delta E (p) = 70 mV, I (pa)/I (pc) = 0.89) were found close to an ideal amperometric transducer. The biosensor on the detection of glucose at 200 mV (vs. Ag/AgCl) showed a linear response range (0-4 mM). The response time of the biosensor was about 10 s. Finally, the biosensor was used together with reverse iontophoresis technique. By the use of an actual model for evaluation, an excellent linear relationship (r (2) = 0.99) was found between the glucose concentration of the actual model and the biosensor current response. In conclusion, a ferrocene mediated glucose biosensor incorporated with reverse iontophoresis function was developed.
引用
收藏
页码:1548 / 1555
页数:8
相关论文
共 58 条
[1]   A MEDIATED AMPEROMETRIC ENZYME ELECTRODE USING TETRATHIAFULVALENE AND L-GLUTAMATE OXIDASE FOR THE DETERMINATION OF L-GLUTAMIC ACID [J].
ALMEIDA, NF ;
MULCHANDANI, AK .
ANALYTICA CHIMICA ACTA, 1993, 282 (02) :353-361
[2]  
ANGERSTEINKOZLO.H, 1984, COMPREHENSIVE TREATI
[3]   STUDIES ON ENZYME ELECTRODES WITH FERROCENE AND CARBON PASTE BOUND WITH CELLULOSE TRIACETATE [J].
BEH, SK ;
MOODY, GJ ;
THOMAS, JDR .
ANALYST, 1991, 116 (05) :459-462
[4]  
Buckingham Bruce, 2005, Diabetes Technol Ther, V7, P440, DOI 10.1089/dia.2005.7.440
[5]  
Cass A.E. G., 1990, Biosensors: A Practical Approach
[6]   FERROCENE-MEDIATED ENZYME ELECTRODE FOR AMPEROMETRIC DETERMINATION OF GLUCOSE [J].
CASS, AEG ;
DAVIS, G ;
FRANCIS, GD ;
HILL, HAO ;
ASTON, WJ ;
HIGGINS, IJ ;
PLOTKIN, EV ;
SCOTT, LDL ;
TURNER, APF .
ANALYTICAL CHEMISTRY, 1984, 56 (04) :667-671
[7]  
Cattrall R.W., 1997, CHEM SENSORS
[8]   Novel diffusion cell ideal for the study of membrane extraction/permeation processes and for device/sensor development [J].
Ching, C. T. S. ;
Connolly, P. .
SENSORS AND ACTUATORS B-CHEMICAL, 2008, 129 (01) :30-34
[9]   A low-cost, programmable device for versatile current delivery in iontophoresis applications [J].
Ching, CTS ;
Camilleri, I ;
Connolly, P .
SENSORS AND ACTUATORS B-CHEMICAL, 2005, 106 (02) :534-540
[10]   Glucose biosensors based on electrodes modified with ferrocene derivatives intercalated into Mg/Al layered double hydroxides [J].
Colombari, M. ;
Ballarin, B. ;
Carpani, I. ;
Guadagnini, L. ;
Mignani, A. ;
Scavetta, E. ;
Tonelli, D. .
ELECTROANALYSIS, 2007, 19 (22) :2321-2327