Integration of a nanoporous platinum thin film into a microfluidic system for non-enzymatic electrochemical glucose sensing

被引:42
作者
Joo, Segyeong
Park, Sejin
Chung, Thek Dong [1 ]
Kim, Hee Chan
机构
[1] Sungshin Womens Univ, Ctr NanoBio Appl Technol, Seoul 136742, South Korea
[2] Seoul Natl Univ, Interdisciplinary Program, Seoul 110744, South Korea
[3] Sungshin Womens Univ, Dept Chem, Seoul 136742, South Korea
[4] Seoul Natl Univ, Coll Med, Dept Biomed Engn, Seoul 110744, South Korea
[5] Seoul Natl Univ, Med Res Ctr, Inst Med & Biol Engn, Seoul 110744, South Korea
关键词
D O I
10.2116/analsci.23.277
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this paper, we describe an amperometric-type enzymeless glucose sensing system based on a nanoporous platinum (Pt) electrode embedded in a microfluidic chip. This microchip system is comprised of a microfluidic transport channel network and a miniaturized electrochemical cell for nonenzymatic glucose sensing. Sample and buffer solutions were transferred to the cell by programmed electroosmotic flow (EOF). A nanoporous Pt electrode with the roughness factor of 200.6 was utilized to determine glucose concentrations in phosphate buffered saline (PBS) by the direct oxidation of glucose, without any separation process. The sensitivity of the developed system is 1.65 mu A cm(-2) mM(-1) in the glucose concentration range from 1 - 10 mM in PBS.
引用
收藏
页码:277 / 281
页数:5
相关论文
共 23 条
[1]   On chip single-cell separation and immobilization using optical tweezers and thermosensitive hydrogel [J].
Arai, F ;
Ng, C ;
Maruyama, H ;
Ichikawa, A ;
El-Shimy, H ;
Fukuda, T .
LAB ON A CHIP, 2005, 5 (12) :1399-1403
[2]   Microstructure for efficient continuous flow mixing [J].
Bessoth, FG ;
deMello, AJ ;
Manz, A .
ANALYTICAL COMMUNICATIONS, 1999, 36 (06) :213-215
[3]   PULSED AMPEROMETRIC DETECTION OF GLUCOSE IN BIOLOGICAL-FLUIDS AT A SURFACE-MODIFIED GOLD ELECTRODE [J].
BINDRA, DS ;
WILSON, GS .
ANALYTICAL CHEMISTRY, 1989, 61 (22) :2566-2570
[4]   Ionic strength-controlled virtual area of mesoporous platinum electrode [J].
Boo, H ;
Park, S ;
Ku, BY ;
Kim, Y ;
Park, JH ;
Kim, HC ;
Chung, TD .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2004, 126 (14) :4524-4525
[5]   Silicon microchambers for DNA amplification [J].
Daniel, JH ;
Iqbal, S ;
Millington, RB ;
Moore, DF ;
Lowe, CR ;
Leslie, DL ;
Lee, MA ;
Pearce, MJ .
SENSORS AND ACTUATORS A-PHYSICAL, 1998, 71 (1-2) :81-88
[6]   MANIPULATION OF SAMPLE FRACTIONS ON A CAPILLARY ELECTROPHORESIS CHIP [J].
EFFENHAUSER, CS ;
MANZ, A ;
WIDMER, HM .
ANALYTICAL CHEMISTRY, 1995, 67 (13) :2284-2287
[7]   Direct determination of carbohydrates, amino acids, and antibiotics by microchip electrophoresis with pulsed amperometric detection [J].
García, CD ;
Henry, CS .
ANALYTICAL CHEMISTRY, 2003, 75 (18) :4778-4783
[8]  
HEEREN FV, 1996, J MICROCOLUMN SEP, V8, P373
[9]   PMMA-based capillary electrophoresis electrochemical detection microchip fabrication [J].
Horng, RH ;
Han, P ;
Chen, HY ;
Lin, KW ;
Tsai, TM ;
Zen, JM .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2005, 15 (01) :6-10
[10]   Power-free poly(dimethylsiloxane) microfluidic devices for gold nanoparticle-based DNA analysis [J].
Hosokawa, K ;
Sato, K ;
Ichikawa, N ;
Maeda, M .
LAB ON A CHIP, 2004, 4 (03) :181-185