Subcutaneous glucose sensor values closely parallel blood glucose during insulin-induced hypoglycaemia

被引:54
作者
Caplin, NJ [1 ]
O'Leary, P [1 ]
Bulsara, M [1 ]
Davis, EA [1 ]
Jones, TW [1 ]
机构
[1] Princess Margaret Hosp Children, Dept Endocrinol & Diabet, Perth, WA, Australia
关键词
hypoglycaemia; glucose sensing; diabetes;
D O I
10.1046/j.1464-5491.2003.00837.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims To assess the accuracy of the Minimed continuous glucose monitoring system (CGMS) in estimating blood glucose concentration during a controlled reduction in blood glucose. Research design and methods We studied nine adolescent diabetics (age 14+/-1.5 years) wearing the CGMS during hyperinsulinaemic hypoglycaemic clamp studies. The glucose values obtained by the CGMS were compared with the venous blood samples taken during the studies and measured at the bedside using a glucose oxidase technique. Results Blood glucose was lowered from euglycaemia to a mean of 2.8 mmol/l over 120 min and maintained at that level for a further 40 min. A total of 429 paired glucose measurements were available for analysis. Analysis using weighted Deming regression and t -tests revealed small differences between the methods, with blood glucose levels slightly higher than interstitial fluid levels. The mean difference across all values was 0.13 mmol/l. The observed difference was greatest at blood glucose values < 3.3 mmol/l. Conclusions This study suggests that during a fall in systemic glucose the subcutaneous glucose sensor provides an accurate reflection of blood glucose. However, a small difference is apparent at blood glucose values < 3.3 mmol/l.
引用
收藏
页码:238 / 241
页数:4
相关论文
共 21 条
[1]   DEVELOPMENT OF A WEARABLE GLUCOSE SENSOR - STUDIES IN HEALTHY-VOLUNTEERS AND IN DIABETIC-PATIENTS [J].
AALDERS, AL ;
SCHMIDT, FJ ;
SCHOONEN, AJM ;
BROEK, IR ;
MAESSEN, AGFM ;
DOORENBOS, H .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1991, 14 (02) :102-108
[2]   Interstitial glucose concentration and glycemia: implications for continuous subcutaneous glucose monitoring [J].
Aussedat, B ;
Dupire-Angel, M ;
Gifford, R ;
Klein, JC ;
Wilson, GS ;
Reach, G .
AMERICAN JOURNAL OF PHYSIOLOGY-ENDOCRINOLOGY AND METABOLISM, 2000, 278 (04) :E716-E728
[3]   Self-monitoring of blood glucose in type I diabetic patients: Comparison with continuous microdialysis measurements of glucose in subcutaneous adipose tissue during ordinary life conditions [J].
Bolinder, J ;
HagstromToft, E ;
Ungerstedt, U ;
Arner, P .
DIABETES CARE, 1997, 20 (01) :64-70
[4]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[5]   SUPPRESSION OF COUNTERREGULATORY HORMONE RESPONSE TO HYPOGLYCEMIA BY INSULIN PER SE [J].
DIAMOND, MP ;
HALLARMAN, L ;
STARICKZYCH, K ;
JONES, TW ;
CONNOLLYHOWARD, M ;
TAMBORLANE, WV ;
SHERWIN, RS .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 1991, 72 (06) :1388-1390
[6]   CLINICAL USEFULNESS OF THE GLUCOSE-CONCENTRATION IN THE SUBCUTANEOUS TISSUE - PROPERTIES AND PITFALLS OF ELECTROCHEMICAL BIOSENSORS [J].
FISCHER, U ;
REBRIN, K ;
VONWOEDTKE, T ;
ABEL, P .
HORMONE AND METABOLIC RESEARCH, 1994, 26 (11) :515-522
[7]   CHARACTERIZATION BY MICRODIALYSIS OF INTERCELLULAR GLUCOSE LEVEL IN SUBCUTANEOUS TISSUE IN HUMANS [J].
JANSSON, PA ;
FOWELIN, J ;
SMITH, U ;
LONNROTH, P .
AMERICAN JOURNAL OF PHYSIOLOGY, 1988, 255 (02) :E218-E220
[8]  
LINNET K, 1993, CLIN CHEM, V39, P424
[9]   ESTIMATION OF THE LINEAR RELATIONSHIP BETWEEN THE MEASUREMENTS OF 2 METHODS WITH PROPORTIONAL ERRORS [J].
LINNET, K .
STATISTICS IN MEDICINE, 1990, 9 (12) :1463-1473
[10]   Microdialysis in adipose tissue and skeletal muscle [J].
Lonnroth, P .
HORMONE AND METABOLIC RESEARCH, 1997, 29 (07) :344-346