Traumatic loading of the Bryan cervical disc prosthesis: An in vitro study

被引:6
作者
Duggal, Neil
Rabin, Doron
Chamberlain, Robert H.
Baek, Seungwon
Crawford, Neil R.
机构
[1] Barrow Neurol Inst, Phoenix, AZ 85013 USA
[2] Univ Western Ontario, Div Neurosurg, London, ON, Canada
关键词
artificial disc; Bryan disc; cervical spine; disc replacement; spinal arthroplasty; trauma;
D O I
10.1227/01.NEU.0000255374.34475.D0
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: The Bryan disc prosthesis (Medtronic Sofamor Danek, Memphis, TN) relies on a precise fit between the device and the vertebral endplates to provide immediate stability after cervical arthroplasty. The safety of the cervical arthroplasty in the setting of trauma is unknown. We compare the segmental strength of the normal cervical spine and the cervical spine after single-level arthroplasty. METHODS: Fifteen cadaveric cervical spines with arthroplasty placed at the C5-C6 level were compared with 16 intact cadaveric controls. A pure moment was applied to induce flexion, extension, or axial rotation until the segment failed. RESULTS: The prosthesis provided 63, 45, and 69% of the strength of a normal spine during flexion, extension, and rotation, respectively. There were no cases of prosthesis expulsion. CONCLUSION: After insertion of the Bryan disc, the remaining ligamentous tissues provide adequate acute stability to the spine.
引用
收藏
页码:388 / 392
页数:5
相关论文
共 18 条
[1]
Biomechanical analysis of a resorbable anterior cervical graft containment plate [J].
Ames, CP ;
Crawford, NR ;
Chamberlain, RH ;
Deshmukh, V ;
Sadikovic, B ;
Sonntag, VKH .
SPINE, 2005, 30 (09) :1031-1038
[2]
A comparison of simulator-tested and -retrieved cervical disc prostheses [J].
Anderson, PA ;
Rouleau, JP ;
Toth, JM ;
Riew, KD .
JOURNAL OF NEUROSURGERY-SPINE, 2004, 1 (02) :202-210
[3]
Anderson Paul A, 2004, Spine J, V4, p303S, DOI 10.1016/j.spinee.2004.07.026
[4]
A new technique for determining 3-D joint angles: the tilt/twist method [J].
Crawford, NR ;
Yamaguchi, GT ;
Dickman, CA .
CLINICAL BIOMECHANICS, 1999, 14 (03) :153-165
[5]
AN APPARATUS FOR APPLYING PURE NONCONSTRAINING MOMENTS TO SPINE SEGMENTS IN-VITRO [J].
CRAWFORD, NR ;
BRANTLEY, AGU ;
DICKMAN, CA ;
KOENEMAN, EJ .
SPINE, 1995, 20 (19) :2097-2100
[6]
Construction of local vertebral coordinate systems using a digitizing probe - Technical note [J].
Crawford, NR ;
Dickman, CA .
SPINE, 1997, 22 (05) :559-563
[7]
Preliminary clinical experience with the Bryan Cervical Disc Prosthesis [J].
Goffin, J ;
Casey, A ;
Kehr, P ;
Liebig, K ;
Lind, B ;
Logroscino, C ;
Pointillart, V ;
Van Calenbergh, F ;
van Loon, J .
NEUROSURGERY, 2002, 51 (03) :840-845
[8]
Intermediate follow-up after treatment of degenerative disc disease with the Bryan Cervical Disc Prosthesis: Single-level and bi-level [J].
Goffin, J ;
Van Calenbergh, F ;
van Loon, J ;
Casey, A ;
Kehr, P ;
Liebig, K ;
Lind, B ;
Logroscino, C ;
Sgrambiglia, R ;
Pointillart, V .
SPINE, 2003, 28 (24) :2673-2678
[9]
Comparison of stiffness and failure load of two cervical spine fixation techniques in an in vitro human model [J].
Hart, R ;
Gillard, J ;
Prem, S ;
Shea, M ;
Kitchel, S .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2005, 18 :S115-S118
[10]
Cervical disc replacement - Porous Coated Motion prosthesis - A comparative biomechanical analysis showing the key role of the posterior longitudinal ligament [J].
McAfee, PC ;
Cunningham, B ;
Dmitriev, A ;
Hu, NB ;
Kim, SW ;
Cappuccino, A ;
Pimenta, L .
SPINE, 2003, 28 (20) :S176-S185