Subfemtomole MS and MS/MS peptide sequence analysis using nano-HPLC micro-ESI Fourier transform ion cyclotron resonance mass spectrometry

被引:281
作者
Martin, SE [1 ]
Shabanowitz, J [1 ]
Hunt, DF [1 ]
Marto, JA [1 ]
机构
[1] Univ Virginia, Dept Chem, Charlottesville, VA 22904 USA
关键词
D O I
10.1021/ac000497v
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Subfemtomole peptide sequence analysis has been achieved using microcapillary HPLC columns, with integrated nanoelectrospray emitters, coupled directly to a Fourier transform ion cyclotron resonance mass spectrometer. Accurate mass (+/-0.010 Da) peptide maps are generated from a standard six-protein digest mixture, whose principle components span a concentration dynamic range of 1000:1. Iterative searches against similar to 189 000 entries in the OWL database readily identify each protein, with high sequence coverage (20-60%), from as little as 10 amol loaded on-column. In addition, a simple variable-flow HPLC apparatus provides for on-line tandem mass spectrometric analysis of tryptic peptides at the 400-amol level. MS/MS data are searched against similar to 280 000 entries in a nonredundant protein database using SEQUEST. Accurate precursor and product ion mass information readily identifiers primary amino acid sequences differing by:asparagine vs aspartic acid (Delta m = 0.98 Da) and glutamine vs lysine (Delta m = 0.036 Da).
引用
收藏
页码:4266 / 4274
页数:9
相关论文
共 84 条