Use of 1.5-μm porous ethyl-bridged hybrid particles as a stationary-phase support for reversed-phase ultrahigh-pressure liquid chromatography

被引:157
作者
Mellors, JS [1 ]
Jorgenson, JW [1 ]
机构
[1] Univ N Carolina, Dept Chem, Chapel Hill, NC 27599 USA
关键词
D O I
10.1021/ac049643d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new ethyl-bridged hybrid packing material was evaluated in terms of its suitability for ultrahigh-pressure liquid chromatography (UHPLC). The 1.5-mum particles were obtained and packed into 30-mum-i.d. fused-silica capillary columns up to 50 cm in length. The particles were evaluated by isocratic reversed-phase UHPLC at pressures up to 4500 bar (65 000 psi). The chromatographic performance of these particles was found to be similar to the performance of 1.0-mum nonporous silica particles. The mechanical strength of the ethyl-bridged hybrid material was evaluated by running a 15-cm-long column at pressures up to 4500 bar. No breakdown of the particles in the packed bed was observed. The sample loading capacity of the hybrid material was evaluated and compared to 1.0-mum nonporous silica material by observing analyte peak width versus amount injected. The observed improvement in loading capacity for the hybrid material versus nonporous silica was consistent with the improvement predicted by comparing the phase ratios of the two materials.
引用
收藏
页码:5441 / 5450
页数:10
相关论文
共 19 条
  • [11] PATEL KD, 2001, THESIS U N CAROLINA
  • [12] Snyder J. J., 1979, INTRO MODERN LIQUID
  • [13] Stellwagen NC, 2001, BIOPOLYMERS, V58, P390, DOI 10.1002/1097-0282(20010405)58:4<390::AID-BIP1015>3.0.CO
  • [14] 2-K
  • [15] Stoeber W, 1968, J COLLOID INTERF SCI, V26, P62
  • [16] WALBROEHL Y, 1989, J MICROCOLUMN SEP, V1, P41
  • [17] Characterization and evaluation of C18HPLC stationary phases based on ethyl-bridged hybrid organic/inorganic particles
    Wyndham, KD
    O'Gara, JE
    Walter, TH
    Glose, KH
    Lawrence, NL
    Alden, BA
    Izzo, GS
    Hudalla, CJ
    Iraneta, PC
    [J]. ANALYTICAL CHEMISTRY, 2003, 75 (24) : 6781 - 6788
  • [18] Safety concerns in ultrahigh pressure capillary liquid chromatography using air-driven pumps
    Xiang, YQ
    Maynes, DR
    Lee, ML
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2003, 991 (02) : 189 - 196
  • [19] DETERMINATION OF DIFFUSION-COEFFICIENTS BY CAPILLARY ZONE ELECTROPHORESIS
    YAO, YJ
    LI, SFY