Chasing the heart - New developments for cardiac CT

被引:38
作者
Flohr, Thomas G.
Schoepf, U. Joseph
Ohnesorge, Bernd M.
机构
[1] Siemens Med Solut, Computed Tomog, D-91301 Forchheim, Germany
[2] Univ Tubingen, Dept Diagnost Radiol, D-72074 Tubingen, Germany
[3] Med Univ S Carolina, Dept Radiol, Charleston, SC 29425 USA
[4] Med Univ S Carolina, Dept Med, Charleston, SC 29425 USA
关键词
computed tomography; dual-source CT technology; heart disease; coronary artery disease; flat panel CT technology;
D O I
10.1097/RTI.0b013e318032132c
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
With the latest generations of multidetector row computed tomography (CT) scanners, CT of the heart is about to fulfill its promise to become the premier noninvasive imaging modality for the cardiac assessment. The performance of this modality has been continuously improved to a point where CT, beyond mere feasibility studies, is firmly establishing its role in the diagnostic work-up of patients with suspected cardiac disease. This has been enabled by ongoing technical refinements, which are the topic of this contribution. This review traces the evolution of CT for cardiac applications, describes the current status of scanner technology with special emphasis on dual-source CT, and provides insights into potential future developments for further refinement of this technique.
引用
收藏
页码:4 / 16
页数:13
相关论文
共 59 条
  • [1] Achenbach S, 2000, CIRCULATION, V102, P2823
  • [2] Contrast-enhanced coronary artery visualization by dual-source computed tomography - Initial experience
    Achenbach, S
    Ropers, D
    Kuettner, A
    Flohr, T
    Ohnesorge, B
    Bruder, H
    Theessen, H
    Karakaya, M
    Daniel, WG
    Bautz, W
    Kalender, WA
    Anders, K
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2006, 57 (03) : 331 - 335
  • [3] Curved multiplanar reconstructions for the evaluation of contrast-enhanced electron beam CT of the coronary arteries
    Achenbach, S
    Moshage, W
    Ropers, D
    Bachmann, K
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1998, 170 (04) : 895 - 899
  • [4] In-plane coronary arterial motion velocity: Measurement with electron-beam CT
    Achenbach, S
    Ropers, D
    Holle, J
    Muschiol, G
    Daniel, WG
    Moshage, W
    [J]. RADIOLOGY, 2000, 216 (02) : 457 - 463
  • [5] QUANTIFICATION OF CORONARY-ARTERY CALCIUM USING ULTRAFAST COMPUTED-TOMOGRAPHY
    AGATSTON, AS
    JANOWITZ, WR
    HILDNER, FJ
    ZUSMER, NR
    VIAMONTE, M
    DETRANO, R
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (04) : 827 - 832
  • [6] BAHNER ML, 1999, EUR RADIOL, V9, P106
  • [7] Imaging of noncalcified coronary plaques using helical CT with retrospective ECG gating
    Becker, CR
    Knez, A
    Ohnesorge, B
    Schoepf, UJ
    Reiser, MF
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 175 (02) : 423 - 424
  • [8] Detection and quantification of coronary artery calcification with electron-beam and conventional CT
    Becker, CR
    Knez, A
    Jakobs, TF
    Aydemir, S
    Becker, A
    Schoepf, UJ
    Bruening, R
    Haberl, R
    Reiser, MF
    [J]. EUROPEAN RADIOLOGY, 1999, 9 (04) : 620 - 624
  • [9] Coronary artery calcium measurement: Agreement of multirow detector and electron beam CP
    Becker, CR
    Kleffel, T
    Crispin, A
    Knez, A
    Young, J
    Schoepf, UJ
    Haberl, R
    Reiser, MF
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (05) : 1295 - 1298
  • [10] Ultrafast computed tomography as a diagnostic modality in the detection of coronary artery disease: A multicenter study
    Budoff, MJ
    Georgiou, D
    Brody, A
    Agatston, AS
    Kennedy, J
    Wolfkiel, C
    Stanford, W
    Shields, P
    Lewis, RJ
    Janowitz, WR
    Rich, S
    Brundage, BH
    [J]. CIRCULATION, 1996, 93 (05) : 898 - 904