Diagnostic x-ray spectra: A comparison of spectra generated by different computational methods with a measured spectrum

被引:57
作者
Bhat, M [1 ]
Pattison, J
Bibbo, G
Caon, M
机构
[1] Univ S Australia, Sch Phys & Elect Syst Engn, The Levels, SA 5095, Australia
[2] Womens & Childrens Hosp, Div Med Imaging, N Adelaide, SA 5006, Australia
[3] Flinders Univ S Australia, Sch Nursing, Bedford Pk, SA 5042, Australia
关键词
semi-empirical model; x-ray spectrum; high-purity germanium detector;
D O I
10.1118/1.598170
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A number of computer codes, developed using semi-empirical models, are available to compute x-ray spectra from a tungsten target for different tube parameters, In this study x-ray spectra measured with a high-purity germanium detector are compared with those computed using the empirical models and previously published measured data. The computer codes used to generate the spectra are based on models proposed by Birch et al. and Tucker et al. The measured x-ray spectra agreed well with the computed x-ray spectra using the model of Tucker et al, whereas the model of Birch et al. produced a "harder" x-ray spectrum compared to the measured spectra. Our measured x-ray spectra compared well with the previously published measured spectral data of Fewell et al. (C) 1998 American Association of Physicists in Medicine.
引用
收藏
页码:114 / 120
页数:7
相关论文
共 13 条
[1]   COMPUTATION OF BREMSSTRAHLUNG X-RAY-SPECTRA AND COMPARISON WITH SPECTRA MEASURED WITH A GE(LI) DETECTOR [J].
BIRCH, R ;
MARSHALL, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (03) :505-517
[2]  
BIRCH R, 1979, SCI REPORT SERIES HO, V30
[3]   A COMPARISON OF SEMIEMPIRICAL MODELS FOR GENERATING TUNGSTEN TARGET X-RAY-SPECTRA [J].
BISSONNETTE, JP ;
SCHREINER, LJ .
MEDICAL PHYSICS, 1992, 19 (03) :579-582
[4]   THE 3 PARAMETER EQUIVALENT SPECTRA AS AN INDEX OF BEAM QUALITY [J].
BOONE, JM .
MEDICAL PHYSICS, 1988, 15 (03) :304-310
[5]  
CAON M, UNPUB MED PHYS
[6]  
CRANLEY K, 1991, 64 I PHYS SCI MED
[7]  
Fewell T., 1981, HDB COMPUTED TOMOGRA
[8]  
FEWELL T R, 1977, Medical Physics (Woodbury), V4, P187, DOI 10.1118/1.594364
[9]  
ILES WJ, 1987, NRPBR204
[10]   DETERMINATION OF X-RAY-SPECTRA AND OF THE SCATTERED COMPONENT UP TO 300 KV [J].
LAITANO, RF ;
PANI, R ;
PELLEGRINI, R .
MEDICAL PHYSICS, 1991, 18 (05) :934-938