Schizophrenic behavior of zones in capillary zone electrophoresis: Explanation of an old problem

被引:18
作者
Gebauer, P
Desiderio, C
Fanali, S
Bocek, P
机构
[1] Acad Sci Czech Republ, Inst Analyt Chem, CZ-61142 Brno, Czech Republic
[2] CNR, Ist Cromatog, Area Ric, I-00016 Monterotondo, Roma, Italy
关键词
migration behavior of zones in CZE; physico-chemical description of zones; fronting and tailing of zones;
D O I
10.1002/elps.1150190517
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Background electrolyte (BGE) systems with two colons are frequently used in capillary zone electrophoresis (CZE), especially in cases where indirect optical detection is employed. This study investigates the behavior of analytes, which possess mobilities intermediate to those of the BGE colons used. Besides the expected behavior, where the analytes provide either tailing or fronting zones, unusual behavior with extraordinary zone broadening is also observed in some cases. The explanation for this effect is that binary BGE systems involve, as a physico-chemical rule, a region where the analytes are forced by one colon to give tailing zones and simultaneously by the other colon to give fronting zones. The result of this "schizophrenic" situation is extraordinary zone broadening and deterioration of the detection record. A series of experiments is presented showing in a telling way the electromigration behavior of the discussed type of zones as well as the ways to remedy the deterioration of the peak shape by a mere slight changing of the quantitative composition of the BGE used.
引用
收藏
页码:701 / 706
页数:6
相关论文
共 13 条
[1]   CAPILLARY ELECTROPHORESIS OF ORGANIC AND INORGANIC CATIONS WITH INDIRECT UV DETECTION [J].
BECK, W ;
ENGELHARDT, H .
CHROMATOGRAPHIA, 1992, 33 (7-8) :313-316
[2]   THEORETICAL AND EXPERIMENTAL ASPECTS OF INDIRECT DETECTION IN CAPILLARY ELECTROPHORESIS [J].
BRUIN, GJM ;
VANASTEN, AC ;
XU, XM ;
POPPE, H .
JOURNAL OF CHROMATOGRAPHY, 1992, 608 (1-2) :97-107
[3]   EFFECT OF MULTIPLE ELECTROLYTE BUFFERS ON PEAK SYMMETRY, RESOLUTION, AND SENSITIVITY IN CAPILLARY ELECTROPHORESIS [J].
BULLOCK, J ;
STRASTERS, J ;
SNIDER, J .
ANALYTICAL CHEMISTRY, 1995, 67 (18) :3246-3252
[4]   System peaks in capillary zone electrophoresis .2. Experimental study of vacancy peaks [J].
Desiderio, C ;
Fanali, S ;
Gebauer, P ;
Bocek, P .
JOURNAL OF CHROMATOGRAPHY A, 1997, 772 (1-2) :81-89
[5]   INDIRECT PHOTOMETRIC DETECTION IN CAPILLARY ZONE ELECTROPHORESIS [J].
FORET, F ;
FANALI, S ;
OSSICINI, L ;
BOCEK, P .
JOURNAL OF CHROMATOGRAPHY, 1989, 470 (01) :299-308
[6]   CAPILLARY ZONE ELECTROPHORESIS OF RARE-EARTH-METALS WITH INDIRECT UV ABSORBENCY DETECTION [J].
FORET, F ;
FANALI, S ;
NARDI, A ;
BOCEK, P .
ELECTROPHORESIS, 1990, 11 (09) :780-783
[7]   Predicting peak symmetry in capillary zone electrophoresis: The concept of the peak shape diagram [J].
Gebauer, P ;
Bocek, P .
ANALYTICAL CHEMISTRY, 1997, 69 (08) :1557-1563
[8]  
HELLER C, 1997, ANAL NUCL ACIDS CAPI, V1, P1
[9]  
JANDIK P, 1992, LC GC EUR, V5, P20
[10]   System peaks in capillary zone electrophoresis .3. Practical rules for predicting the existence of system peaks in capillary zone electrophoresis of anions using indirect spectrophotometric detection [J].
Macka, M ;
Haddad, PR ;
Gebauer, P ;
Bocek, P .
ELECTROPHORESIS, 1997, 18 (11) :1998-2007