Early failure of bioabsorbable anterior cervical fusion plates - Case report and failure analysis

被引:19
作者
Brkaric, Mario
Baker, Kevin C.
Israel, Raj
Harding, Trevor
Montgomery, David M.
Herkowitz, Harry N.
机构
[1] William Beaumont Hosp Res Inst, Harold W Gehring MD Ctr Biomech & Implant Retriev, Royal Oak, MI 48073 USA
[2] Kettering Univ, Dept Ind & Mfg Engn, Flint, MI USA
[3] William Beaumont Hosp, Dept Orthoped Surg, Royal Oak, MI 48072 USA
来源
JOURNAL OF SPINAL DISORDERS & TECHNIQUES | 2007年 / 20卷 / 03期
关键词
failure analysis; bioabsorbable polymers; cervical spine fusion; retrieval study;
D O I
10.1097/BSD.0b013e318030d2f3
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design: Case report with forensic failure analysis. Objective: To determine the failure modes of 3 explanted 70:30 PLDLA Mystique (Medtronic Sofamor Danek, Memphis, TN) graft containment plates retrieved from revision surgery for early device failure. Summary of Background Data: To reduce the problems of stress-shielding and radiopacity associated with metallic systems, bioabsorbable polymers have been used in anterior cervical discectomy and fusion procedures. Degradation of mechanical properties in vivo is a major concern when using bioabsorbable systems. Three of 6 patients who underwent anterior cervical discectomy with instrumented fusion, using Mystique graft containment systems experienced early failure requiring revision to alternate hardware. Methods: Devices were retrieved after failure and analyzed by light microscopy and environmental scanning electron microscopy. Simulations were performed with an unused plating system to induce damage for comparison with the retrieved devices. A detailed case review was performed to identify possible sources of extraordinary loading or damage. Results: One plating system failed at 6 weeks postimplantation due to fatigue fracture of the screws. Crack initiation sites were identified at the interface of the thread root and mold line of the screw. Another plating system failed at 16 weeks postimplantation due to the coalescence of radial microcracking between holes in the plate, leading to catastrophic failure of the plate. The final plating system failed during the implantation surgery, when the screw fractured in torsion. Conclusions: Stress concentrations at the screw head-shaft interface and thread-shaft interface reduce the fatigue performance of bioabsorbable screws. Hydrolysis of the polymer may also play a role in the reduction of resistance to crack initiation and propagation.
引用
收藏
页码:248 / 254
页数:7
相关论文
共 26 条
[1]
ANDERSON TL, 1995, FRACTURE MECHANICS
[2]
*ASM INT, 1987, ASH HDB FRACT, V12
[3]
Callister WD., 2015, FUNDAMENTALS MAT SCI, DOI 10.1007/978-981-10-2529-7_2
[4]
Osteollysis caused by tibial component debonding in total knee arthroplasty [J].
Cheng, Kevin ;
Pruitt, Lisa ;
Zaloudek, Charles ;
Ries, Michael D. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (443) :333-336
[5]
Resorbable polymer fibers for ligament augmentation [J].
Dürselen, L ;
Dauner, M ;
Hierlemann, H ;
Planck, H ;
Claes, LE ;
Ignatius, A .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2001, 58 (06) :666-672
[6]
Eppley BL, 2004, PLAST RECONSTR SURG, V114, P850, DOI 10.1097/01.PRS.0000132856.69391.43
[7]
Treatment of syndesmotic disruptions of the ankle with bioabsorbable screw fixation [J].
Hovis, WD ;
Kaiser, BW ;
Watson, JT ;
Bucholz, RW .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (01) :26-31
[8]
Anterior cervical plating enhances arthrodesis after discectomy and fusion with cortical allograft [J].
Kaiser, MG ;
Haid, RW ;
Subach, BR ;
Barnes, B ;
Rodts, GE .
NEUROSURGERY, 2002, 50 (02) :229-236
[9]
Meniscal surgery 2002 update: Indications and techniques for resection, repair, regeneration, and replacement [J].
Klimkiewicz, JJ ;
Shaffer, B .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2002, 18 (09) :14-25
[10]
Treatment of pectus excavatum with bioabsorbable polylactide plates:: Preliminary results [J].
Länsman, S ;
Serlo, W ;
Linna, O ;
Pohjonen, T ;
Törmälä, P ;
Waris, T ;
Ashammakhi, N .
JOURNAL OF PEDIATRIC SURGERY, 2002, 37 (09) :1281-1286