Chronic renal blood flow measurement in dogs by transit-time ultrasound flowmetry

被引:8
作者
Evans, RG [1 ]
Stevenson, KM [1 ]
Malpas, SC [1 ]
Fitzgerald, SM [1 ]
Shweta, A [1 ]
Tomoda, F [1 ]
Anderson, WP [1 ]
机构
[1] BAKER MED RES INST,PRAHRAN,VIC 3181,AUSTRALIA
关键词
calibration; dog; effective renal plasma flow; para-aminohippuric acid clearance; renal blood flow; transit-time; ultrasound flowmetry;
D O I
10.1016/S1056-8719(97)00046-4
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
To test the validity of transit-time ultrasound flowmetry for chronic measurement of renal blood flow in dogs, we compared this method with the renal clearance of para-aminohippuric acid (C-PAH) (corrected for hematocrit), and with direct volumetric measurements. When flow-probes were implanted without silastic sheeting to stabilize the implant, there was significant disparity between the (within-dog) mean levels of renal blood flow estimated by flow-probe and C-PAH. In contrast, when the flow-probe implants were stabilized with silicone sheeting, there was close agreement in each dog between the flow rates measured by the two methods. When flow-probes were calibrated volumetrically in situ, there was a close linear relationship between flow derived from the flow-probe and that measured volumetrically (r = 0.98 +/- 0.02). We conclude that valid, chronic measurement of renal blood flow in dogs can be achieved using transit-time ultrasound flowmetry, provided the implant is stabilized with silicone sheeting. (C) 1997 Elsevier Science Inc.
引用
收藏
页码:33 / 39
页数:7
相关论文
共 32 条
[1]  
[Anonymous], 1978, P SAN DIEG BIOM S
[2]  
*AUSTR GOV PUBL SE, 1990, AUSTR COD PRACT CAR
[3]   ROLE OF ENDOTHELIUM-DERIVED RELAXING FACTOR IN RENAL AUTOREGULATION IN CONSCIOUS DOGS [J].
BAUMANN, JE ;
PERSSON, PB ;
EHMKE, H ;
NAFZ, B ;
KIRCHHEIM, HR .
AMERICAN JOURNAL OF PHYSIOLOGY, 1992, 263 (02) :F208-F213
[4]   EVALUATION OF A TRANSIT-TIME SYSTEM FOR THE CHRONIC MEASUREMENT OF BLOOD-FLOW IN CONSCIOUS SHEEP [J].
BEDNARIK, JA ;
MAY, CN .
JOURNAL OF APPLIED PHYSIOLOGY, 1995, 78 (02) :524-530
[5]  
BROOKS DP, 1990, J PHARMACOL EXP THER, V254, P1084
[6]   VALIDATION OF 1-MM AND 2-MM TRANSIT-TIME ULTRASOUND FLOW PROBES ON MESENTERIC-ARTERY AND AORTA OF RATS [J].
DALMEIDA, MS ;
GAUDIN, C ;
LEBREC, D .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1995, 268 (03) :H1368-H1372
[7]   NEUROGENIC CONTROL OF PRESSURE NATRIURESIS IN CONSCIOUS DOGS [J].
EHMKE, H ;
PERSSON, PB ;
SEYFARTH, M ;
KIRCHHEIM, HR .
AMERICAN JOURNAL OF PHYSIOLOGY, 1990, 259 (03) :F466-F473
[8]   INTERACTIONS OF BLOCKADE OF NITRIC-OXIDE SYNTHASE AND ANGIOTENSIN-CONVERTING ENZYME ON RENAL-FUNCTION IN CONSCIOUS RABBITS [J].
EVANS, RG ;
RANKIN, AJ ;
ANDERSON, WP .
JOURNAL OF CARDIOVASCULAR PHARMACOLOGY, 1994, 24 (04) :542-551
[9]   RENAL EFFECTS OF INFUSION OF RILMENIDINE AND GUANABENZ IN CONSCIOUS DOGS - CONTRIBUTION OF PERIPHERAL AND CENTRAL-NERVOUS-SYSTEM ALPHA(2)-ADRENOCEPTORS [J].
EVANS, RG ;
ANDERSON, WP .
BRITISH JOURNAL OF PHARMACOLOGY, 1995, 116 (01) :1557-1570
[10]   EFFECTS OF ALPHA-ADRENOCEPTOR ANTAGONISTS AND CLONIDINE ON THE HEMODYNAMIC-RESPONSE TO ACUTE HYPOVOLEMIA IN CONSCIOUS RABBITS [J].
EVANS, RG ;
HAYES, IP ;
LUDBROOK, J .
EUROPEAN JOURNAL OF PHARMACOLOGY, 1992, 216 (02) :265-272