THE COUNTRATE PERFORMANCE OF THE VOLUME IMAGING PENN-PET SCANNER

被引:20
作者
SMITH, RJ [1 ]
KARP, JS [1 ]
MUEHLLEHNER, G [1 ]
机构
[1] UGM MED SYST,PHILADELPHIA,PA 19104
关键词
D O I
10.1109/42.363107
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The UGM PENN-PET Camera uses large position sensitive detectors and operates without septa. This design results in high sensitivity and 3-D imaging capability, but poses problems in high countrate situations. The maximum true countrates and random countrates have been measured, as a function of object size in the field-of-view. The countrate performance is understood in terms of the limiting process rates and event rejection in the camera. In addition, the camera is calibrated to generate absolute activity concentrations to within 5% by correcting for system deadtime with up to 3 mCi in the field of view in brain studies (50% deadtime at 3 mu/mL). This allows the performance of a variety of brain and body studies, and accurate quantitation of the data over a wide range of imaging countrates, from F-18-FDG brain studies to O-15-water dynamic brain studies.
引用
收藏
页码:610 / 618
页数:9
相关论文
共 9 条
[1]  
BAILEY DL, 1991, IEEE T MED IMAGING, V10, P236
[2]  
GERMANO G, 1990, IEEE T NUCL SCI, V27, P671
[3]  
KARP JS, 1990, J NUCL MED, V31, P617
[4]   EFFECT OF INCREASED AXIAL-FIELD OF VIEW ON THE PERFORMANCE OF A VOLUME PET SCANNER [J].
KARP, JS ;
KINAHAN, PE ;
MUEHLLEHNER, G ;
COUNTRYMAN, P .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1993, 12 (02) :299-306
[5]   COMBINATION OF DYNAMIC AND INTEGRAL METHODS FOR GENERATING REPRODUCIBLE FUNCTIONAL CBF IMAGES [J].
LAMMERTSMA, AA ;
CUNNINGHAM, VJ ;
DEIBER, MP ;
HEATHER, JD ;
BLOOMFIELD, PM ;
NUTT, J ;
FRACKOWIAK, RSJ ;
JONES, T .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1990, 10 (05) :675-686
[6]   A LOCAL COINCIDENCE TRIGGERING SYSTEM FOR PET TOMOGRAPHS COMPOSED OF LARGE-AREA POSITION-SENSITIVE DETECTORS [J].
MANKOFF, DA ;
MUEHLLEHNER, G ;
MILES, GE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :730-736
[7]   THE HIGH COUNT RATE PERFORMANCE OF A 2-DIMENSIONALLY POSITION-SENSITIVE DETECTOR FOR POSITRON EMISSION TOMOGRAPHY [J].
MANKOFF, DA ;
MUEHLLEHNER, G ;
KARP, JS .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (04) :437-456
[8]   PHYSICAL PERFORMANCE OF A POSITRON TOMOGRAPH FOR BRAIN IMAGING WITH RETRACTABLE SEPTA [J].
SPINKS, TJ ;
JONES, T ;
BAILEY, DL ;
TOWNSEND, DW ;
GROOTOONK, S ;
BLOOMFIELD, PM ;
GILARDI, MC ;
CASEY, ME ;
SIPE, B ;
REED, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (08) :1637-1655
[9]   MEASURING PET SCANNER SENSITIVITY - RELATING COUNTRATES TO IMAGE SIGNAL-TO-NOISE RATIOS USING NOISE EQUIVALENT COUNTS [J].
STROTHER, SC ;
CASEY, ME ;
HOFFMAN, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :783-788