EFFECT OF INCREASED AXIAL-FIELD OF VIEW ON THE PERFORMANCE OF A VOLUME PET SCANNER

被引:35
作者
KARP, JS [1 ]
KINAHAN, PE [1 ]
MUEHLLEHNER, G [1 ]
COUNTRYMAN, P [1 ]
机构
[1] UGM MED SYST INC,PHILADELPHIA,PA 19104
关键词
D O I
10.1109/42.232259
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The performance of the PENN-PET 240H scanner from UGM Medical Systems is tested and compared to the prototype PENN-PET scanner built at the University of Pennsylvania. The UGM PENN-PET scanner consists of six continuous position-sensitive NaI(Tl) detectors, which results in a 50 cm transverse field-of-view and a 12.8 cm axial field-of-view. The fine spatial sampling in the axial direction allows the data to be sorted into as many as 64 transverse planes, each 2 mm thick. A large axial acceptance angle, without interplane septa, results in a high-sensitivity and low-randoms fraction, with a low-scatter fraction due to the use of a narrow photopeak energy window. This paper emphasizes those performance measurements that illustrate the special characteristics of a volume imaging scanner and how they change as the axial length is increased.
引用
收藏
页码:299 / 306
页数:8
相关论文
共 13 条
  • [1] BAILEY DL, 1991, IEEE T MED IMAGING, V10, P236
  • [2] BURNHAM CA, 1988, IEEE T NUCL SCI, V35, P676
  • [3] DAUBEWITHERSPOON ME, 1987, J NUCL MED, V28, P1717
  • [4] HIGH-RESOLUTION PROPORTIONAL CHAMBER POSITRON CAMERA AND ITS APPLICATIONS
    JEAVONS, AP
    TOWNSEND, DW
    FORD, NL
    KULL, K
    MANUEL, A
    FISCHER, O
    PETER, M
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1978, 25 (01) : 164 - 173
  • [5] KARP JS, 1990, J NUCL MED, V31, P617
  • [6] KARP JS, 1991, J NUCL MED, V32, P2342
  • [7] POSITRON EMISSION TOMOGRAPHY WITH A LARGE AXIAL ACCEPTANCE ANGLE - SIGNAL-TO-NOISE CONSIDERATIONS
    KARP, JS
    KINAHAN, PE
    MANKOFF, DA
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (03) : 249 - 255
  • [8] KINAHAN PE, 1990, P INT C IEEE ENG MED, V12, P383
  • [9] LEWITT RM, UNPUB PHYS MED BIOL
  • [10] A LOCAL COINCIDENCE TRIGGERING SYSTEM FOR PET TOMOGRAPHS COMPOSED OF LARGE-AREA POSITION-SENSITIVE DETECTORS
    MANKOFF, DA
    MUEHLLEHNER, G
    MILES, GE
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) : 730 - 736