2ND-HARMONIC ULTRASONIC BLOOD PERFUSION MEASUREMENT

被引:193
作者
SCHROPE, BA [1 ]
NEWHOUSE, VL [1 ]
机构
[1] DREXEL UNIV,INST BIOMED ENG & SCI,PHILADELPHIA,PA 19104
基金
美国国家科学基金会;
关键词
BLOOD PERFUSION MEASUREMENT; DOPPLER; 2ND HARMONIC; ULTRASONIC CONTRAST AGENT;
D O I
10.1016/0301-5629(93)90080-8
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
In vitro and in vivo testing of a recently introduced method of evaluating blood perfusion is presented, where the Doppler shift of the second harmonic component of the backscattered echo is measured. Central to this measurement is the administration of a galactose-based contrast agent (Schering AG, Berlin, Germany, SHU-508 or derivative) which has been shown in vitro to exhibit extraordinary nonlinear backscattering properties. Two types of experiments are described: in vitro studies on excised sheep kidneys and in vivo studies on living rabbits. In the animal model, blood perfusion was manipulated by various mechanisms to obtain some indication of the quantitative ability of the measurement. Comparisons between measurements made at the fundamental component of the backscattered echo and at the second harmonic show that use of the second harmonic measurement results in a much improved ratio of blood echo intensity to tissue echo intensity (signal-to-clutter ratio), allowing detection of blood flowing in smaller vessels and opening up the potential for real-time determination of blood volume fluctuations in tissue.
引用
收藏
页码:567 / 579
页数:13
相关论文
共 7 条
[1]  
Bjorno L., 1986, TISSUE CHARACTERIZAT, P141
[2]   DESIGN FOR AN ULTRASOUND-BASED INSTRUMENT FOR MEASUREMENT OF TISSUE BLOOD-FLOW [J].
BURNS, SM ;
REID, MH .
BIOMATERIALS ARTIFICIAL CELLS AND ARTIFICIAL ORGANS, 1989, 17 (01) :61-68
[3]   MEASUREMENT OF BLOOD PERFUSION IN TISSUE USING DOPPLER ULTRASOUND [J].
DYMLING, SO ;
PERSSON, HW ;
HERTZ, CH .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (05) :433-444
[4]  
RUCKEBUSH Y, 1991, PHYSL SMALL LARGE AN
[5]  
SAUMET JL, 1986, INT J MICROCIRC, V5, P73
[6]   SIMULATED CAPILLARY BLOOD-FLOW MEASUREMENT USING A NONLINEAR ULTRASONIC CONTRAST AGENT [J].
SCHROPE, B ;
NEWHOUSE, VL ;
UHLENDORF, V .
ULTRASONIC IMAGING, 1992, 14 (02) :134-158
[7]  
TING TH, 1991, ULTRASONIC IMAGING, V13, P210