Molecular weight determination of megadalton DNA electrospray ions using charge detection time-of-flight mass spectrometry

被引:188
作者
Fuerstenau, SD [1 ]
Benner, WH [1 ]
机构
[1] UNIV CALIF BERKELEY,LAWRENCE BERKELEY LAB,DIV ENGN,HUMAN GENOME GRP,BERKELEY,CA 94720
关键词
D O I
10.1002/rcm.1290091513
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We present results obtained with a novel mass spectrometer capable of determining the mass of multiply charged electrospray ions generated from samples of macromolecules in the megadalton (MDa) size range. The instrument utilizes a sensitive amplifier which can detect the charge on a single ion as It passes through a tube detector. A velocity measurement of an ion with known electrostatic energy provides the ion's mass-to-charge ratio. Simultaneous detection of the ion charge permits a mass assignment to be made for each ion. Electrospray ions of DNA and polymer molecules with masses greater than 1 X 10(6) Da and charge numbers (z) in excess of 425 e(-) are readily detected in this mass spectrometer. The weights of small particles were also measured, The on-axis single-ion detection configuration provides a duty cycle of nearly 100% and extends the practical application of electrospray mass spectrometry to the analysis of very large molecules with relatively inexpensive instrumentation.
引用
收藏
页码:1528 / 1538
页数:11
相关论文
共 30 条
[21]  
ROSELLE J, 1994, THESIS YALE U
[22]   ELECTROSTATIC ACCELERATION OF MICROPARTICLES TO HYPERVELOCITIES [J].
SHELTON, H ;
HENDRICKS, CD ;
WUERKER, RF .
JOURNAL OF APPLIED PHYSICS, 1960, 31 (07) :1243-1246
[23]  
SIUZDAK G, 1995, 43RD P ANN C MASS SP
[24]   ULTRA-HIGH VELOCITY IMPACTS - CRATERING STUDIES OF MICROSCOPIC IMPACTS FROM 3-KM/S TO 30-KM/S [J].
STRADLING, GL ;
IDZOREK, GC ;
SHAFER, BP ;
CURLING, HL ;
COLLOPY, MT ;
BLOSSOM, AAH ;
FUERSTENAU, S .
INTERNATIONAL JOURNAL OF IMPACT ENGINEERING, 1993, 14 (1-4) :719-727
[25]   SEPARATION OF LARGE DNA FRAGMENTS BY CAPILLARY ELECTROPHORESIS UNDER PULSED-FIELD CONDITIONS [J].
SUDOR, J ;
NOVOTNY, MV .
ANALYTICAL CHEMISTRY, 1994, 66 (15) :2446-2450
[26]   AN INSTRUMENT FOR MEASURING ELECTRIC CHARGE ON INDIVIDUAL AEROSOL-PARTICLES [J].
VERCOULEN, PHW ;
ROOS, RA ;
MARIJNISSEN, JCM ;
SCARLETT, B .
JOURNAL OF AEROSOL SCIENCE, 1991, 22 :S335-S338
[27]  
Weinheimer A. J., 1988, Journal of Atmospheric and Oceanic Technology, V5, P298, DOI 10.1175/1520-0426(1988)005<0298:TCIOAC>2.0.CO
[28]  
2
[29]   ELECTROSPRAY INTERFACE FOR LIQUID CHROMATOGRAPHS AND MASS SPECTROMETERS [J].
WHITEHOUSE, CM ;
DREYER, RN ;
YAMASHITA, M ;
FENN, JB .
ANALYTICAL CHEMISTRY, 1985, 57 (03) :675-679
[30]   SECONDARY-ELECTRON ANALYSIS OF POLYMERIC IONS GENERATED BY AN ELECTROSPRAY ION-SOURCE [J].
XU, Y ;
BAE, YK ;
BEUHLER, RJ ;
FRIEDMAN, L .
JOURNAL OF PHYSICAL CHEMISTRY, 1993, 97 (46) :11883-11886