HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF AMOXICILLIN IN URINE USING SOLID-PHASE, ION-PAIR EXTRACTION AND ULTRAVIOLET DETECTION

被引:21
作者
CHULAVATNATOL, S [1 ]
CHARLES, BG [1 ]
机构
[1] UNIV QUEENSLAND,DEPT PHARM,BRISBANE,QLD 4072,AUSTRALIA
来源
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS | 1993年 / 615卷 / 01期
关键词
D O I
10.1016/0378-4347(93)80294-E
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple, robust high-performance liquid chromatographic method is described for assaying amoxicillin in urine. Sample clean-up involved solid-phase, ion-pair extraction onto Sep-Pak C18 cartridges followed by elution with acetonitrile (3%, v/v) in pH 4.85 buffer. Separations were performed on an Ultrasphere C18 column with a mobile phase comprising acetonitrile (32.5 ml) and pH 7.1 phosphate buffer (0.01 M, 1000 ml). Peaks were detected at 229 nm. Recovery was greater than 94%. Plots of peak area against urinary amoxicillin concentration were linear (r > 0.999) from 5 to 500 mg/l. Between-day and within-day imprecision (coefficient of variation) ranged between 3.0 and 15.5%. Inaccuracy was 7.1%, or less.
引用
收藏
页码:91 / 96
页数:6
相关论文
共 21 条