The factors governing the choice of preamplifier type for the vibrating capacitive probe used in contact potential measurements are examined. Two types are compared: a high input impedance voltage amplifier and a current amplifier. The latter has been increasingly used in recent years due to its great advantages in dealing with parasitic input capacitance. We extend previous analyses, elucidating other advantages of the current amplifier. Particularly important are (i) the reduction of spurious microphonic signals, implying lower systematic error, and (ii) the white noise spectrum of its equivalent contact potential noise, which allows random error to be effectively reduced by increased averaging periods.
引用
收藏
页码:3744 / 3751
页数:8
相关论文
共 27 条
  • [1] ASHCROFT NW, 1976, SOLID STATE PHYS, pCH18
  • [2] ANALYSIS OF STRAY CAPACITANCE IN THE KELVIN METHOD
    BAIKIE, ID
    VENDERBOSCH, E
    MEYER, JA
    ESTRUP, PJZ
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (03) : 725 - 735
  • [3] ROLE OF CURRENT-AMPLIFIER ASSOCIATED WITH VIBRATING CAPACITOR IN KELVIN METHOD
    BONNET, J
    PALAU, JM
    SOONCKINDT, L
    LASSABATERE, L
    [J]. JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1977, 10 (03): : 212 - 213
  • [4] Craig P. P., 1970, Review of Scientific Instruments, V41, P258, DOI 10.1063/1.1684484
  • [5] EFFECTS OF STRAY CAPACITANCE ON KELVIN METHOD FOR MEASURING CONTACT POTENTIAL DIFFERENCE
    DARCY, RJ
    SURPLICE, NA
    [J]. JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1970, 3 (04) : 482 - &
  • [6] ANALYSIS AND IMPROVEMENT OF KELVIN METHOD FOR MEASURING DIFFERENCES IN WORK FUNCTION
    DEBOER, JSW
    KRUSEMEY.HJ
    JASPERS, NCB
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1973, 44 (08) : 1003 - 1008
  • [7] ACCURATE AND VERSATILE VIBRATING CAPACITOR FOR SURFACE AND ADSORPTION STUDIES
    ENGELHARDT, HA
    FEULNER, P
    PFNUR, H
    MENZEL, D
    [J]. JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1977, 10 (11): : 1133 - 1136
  • [8] AN IMPROVED APPARATUS FOR SURFACE PHOTOVOLTAGE STUDIES WITH A BIMORPHOUS PIEZOELECTRIC KELVIN PROBE
    GERMANOVA, K
    HARDALOV, C
    STRASHILOV, V
    GEORGIEV, B
    [J]. JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1987, 20 (03): : 273 - 276
  • [9] HOROWITZ P, 1980, ART ELECTRONICS
  • [10] Keithley J. E., 2014, LOW LEVEL MEASUREMEN, P1