EX-VIVO EVALUATION OF FERROMAGNETISM AND ARTIFACTS OF CARDIAC OCCLUDERS EXPOSED TO A 1.5-T MR SYSTEM

被引:23
作者
SHELLOCK, FG
MORISOLI, SM
机构
[1] ST JOHNS HOSP, DEPT MED IMAGING, SANTA MONICA, CA USA
[2] UNIV CALIF LOS ANGELES, MED CTR, DEPT RADIOL SCI, LOS ANGELES, CA USA
关键词
ARTIFACT; HEART; PROSTHESES; SAFETY;
D O I
10.1002/jmri.1880040220
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Magnetic resonance (MR) procedures are contraindicated for patients with certain ferromagnetic biomedical implants, primarily owing to the risk of movement or dislodgement of the implants by the static magnetic field. An additional concern is the amount of artifact that the implant produces, which can affect image quality and interpretation of the examination. Therefore, an ex vivo assessment of ferromagnetism and artifact was conducted for 12 different occluders used to treat patients with patent ductus arteriosus, atrial septal defects, and ventricular septal defects, in a 1.5-T MR system. Seven of the occluders, made of 304 stainless steel, were ferromagnetic and displayed deflection forces of 248-299 dynes. Five of the implants, made of MP35n, were nonferromagnetic. Artifacts were variable and depended primarily on the type and amount of metal used to construct the implant. The authors conclude that patients with ferromagnetic cardiac occluders may undergo MR procedures approximately 6 weeks after placement of these devices, to allow tissue growth to provide additional retentive force. After this time, it is highly unlikely that the magnetic fields associated with a 1.5-T MR system are capable of moving or dislodging any of these cardiac occluders.
引用
收藏
页码:213 / 215
页数:3
相关论文
共 20 条
  • [1] IMPORTANT CONSIDERATIONS IN MEASUREMENT OF ATTRACTIVE FORCE ON METALLIC IMPLANTS IN MR IMAGERS
    KAGETSU, NJ
    LITT, AW
    [J]. RADIOLOGY, 1991, 179 (02) : 505 - 508
  • [2] TRANSCATHETER UMBRELLA CLOSURE OF CONGENITAL HEART-DEFECTS
    LOCK, JE
    COCKERHAM, JT
    KEANE, JF
    FINLEY, JP
    WAKELY, PE
    FELLOWS, KE
    [J]. CIRCULATION, 1987, 75 (03) : 593 - 599
  • [3] POTENTIAL HAZARDS AND ARTIFACTS OF FERROMAGNETIC AND NONFERROMAGNETIC SURGICAL AND DENTAL MATERIALS AND DEVICES IN NUCLEAR MAGNETIC-RESONANCE IMAGING
    NEW, PFJ
    ROSEN, BR
    BRADY, TJ
    BUONANNO, FS
    KISTLER, JP
    BURT, CT
    HINSHAW, WS
    NEWHOUSE, JH
    POHOST, GM
    TAVERAS, JM
    [J]. RADIOLOGY, 1983, 147 (01) : 139 - 148
  • [4] NONSURGICAL CLOSURE OF PATENT DUCTUS-ARTERIOSUS - CLINICAL-APPLICATION OF THE RASHKIND PDA OCCLUDER SYSTEM
    RASHKIND, WJ
    MULLINS, CE
    HELLENBRAND, WE
    TAIT, MA
    [J]. CIRCULATION, 1987, 75 (03) : 583 - 592
  • [5] ROME JJ, 1990, CIRCULATION, V80, P753
  • [6] SHELLOCK F G, 1989, Magnetic Resonance Quarterly, V5, P243
  • [7] MR IMAGING AND METALLIC IMPLANTS FOR ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION - ASSESSMENT OF FERROMAGNETISM AND ARTIFACT
    SHELLOCK, FG
    MINK, JH
    CURTIN, S
    FRIEDMAN, MJ
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (02): : 225 - 228
  • [8] MR IMAGING AND BIOMEDICAL IMPLANTS, MATERIALS, AND DEVICES - AN UPDATED REVIEW
    SHELLOCK, FG
    CURTIS, JS
    [J]. RADIOLOGY, 1991, 180 (02) : 541 - 550
  • [9] MR IMAGING OF METALLIC IMPLANTS AND MATERIALS - A COMPILATION OF THE LITERATURE
    SHELLOCK, FG
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (04) : 811 - 814
  • [10] HIGH-FIELD-STRENGTH MR IMAGING AND METALLIC BIOMEDICAL IMPLANTS - AN EXVIVO EVALUATION OF DEFLECTION FORCES
    SHELLOCK, FG
    CRUES, JV
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (02) : 389 - 392