AUTOMATIC IMPLEMENTATION OF THE METHOD OF STANDARD ADDITIONS IN UNSEGMENTED FLOW SYSTEMS

被引:12
作者
AGUDO, M [1 ]
RIOS, A [1 ]
VALCARCEL, M [1 ]
机构
[1] UNIV CORDOBA,FAC SCI,DEPT ANALYT CHEM,E-14004 CORDOBA,SPAIN
关键词
FLOW INJECTION; CALIBRATION; METHOD OF STANDARD ADDITIONS; GLUCOSE; FRUCTOSE;
D O I
10.1016/0003-2670(94)00411-E
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An open/closed flow system allowing variable volumes of both sample and standard calibration solution to be introduced was employed to perform automatic calibration in unsegmented flow systems by using the method of standard additions. The dilution loop thus established was used to homogenize the sample and standard mixture in a constant final volume that was subsequently introduced into the main flow system. The procedure allows direct analysis of liquid samples with no pretreatment and avoids matrix effects. Glucose and fructose were enzymatically determined over the ranges of 0.01-0.1 g l(-1) and 0.02-0.1 g l(-1) with relative standard deviation of +/-0.8% and +/-1.0% for the determination of 0.05 g l(-1) (glucose) and 0.07 g l(-1) (fructose), respectively. Both analytes were determined in various types of drinks with average errors of 3.8% (glucose) and 4.6% (fructose).
引用
收藏
页码:77 / 84
页数:8
相关论文
共 17 条
[1]   AUTOMATIC CALIBRATION AND DILUTION IN UNSEGMENTED FLOW SYSTEMS [J].
AGUDO, M ;
RIOS, A ;
VALCARCEL, M .
ANALYTICA CHIMICA ACTA, 1992, 264 (02) :265-273
[2]   A FAST PROCEDURE FOR STANDARD ADDITIONS IN FLOW-INJECTION ANALYSIS [J].
ARAUJO, MCU ;
PASQUINI, C ;
BRUNS, RE ;
ZAGATTO, EAG .
ANALYTICA CHIMICA ACTA, 1985, 171 (MAY) :337-343
[3]  
BAKER GL, 1984, AUTOMATED STREAM ANA, V2, P36
[4]   NOVEL SINGLE STANDARD CALIBRATION AND DILUTION METHOD PERFORMED BY THE SEQUENTIAL INJECTION TECHNIQUE [J].
BARON, A ;
GUZMAN, M ;
RUZICKA, J ;
CHRISTIAN, GD .
ANALYST, 1992, 117 (12) :1839-1844
[5]   PERMEATION TUBES FOR CALIBRATION IN FLOW-INJECTION ANALYSIS [J].
CHALK, SJ ;
TYSON, JF ;
OLSON, DC .
ANALYST, 1993, 118 (09) :1227-1231
[6]   SEQUENTIAL DETERMINATION OF GLUCOSE AND FRUCTOSE IN FOODS BY FLOW-INJECTION ANALYSIS WITH IMMOBILIZED ENZYMES [J].
LINARES, P ;
DECASTRO, MDL ;
VALCARCEL, M .
ANALYTICA CHIMICA ACTA, 1987, 202 :199-205
[7]   SPECTROPHOTOMETRIC FLOW-INJECTION TECHNIQUES FOR THE MULTICOMPONENT MONITORING OF PROCESS STREAMS [J].
MACLAURIN, P ;
WORSFOLD, PJ .
MICROCHEMICAL JOURNAL, 1992, 45 (02) :178-188
[8]   FLOW-INJECTION DETERMINATION OF SULFITE AND ASSAY OF SULFITE OXIDASE [J].
MASOOM, M ;
TOWNSHEND, A .
ANALYTICA CHIMICA ACTA, 1986, 179 :399-405
[9]   A NEW POSSIBILITY FOR THE CALIBRATION OF POTENTIOMETRIC DETERMINATIONS UNDER FLOW-INJECTION CONDITIONS (FIA) .1. THEORETICAL FUNDAMENTALS [J].
MULLER, H ;
KRAMER, J .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1989, 335 (02) :205-209
[10]   MULTIDETECTION IN UNSEGMENTED FLOW SYSTEMS WITH A SINGLE DETECTOR [J].