ENERGY RESPONSE OF AGAR-ALANINE FREE-RADICAL DOSIMETRY TO THERAPEUTIC ELECTRON-BEAMS

被引:10
作者
CIESIELSKI, B
REINSTEIN, LE
MEEK, AG
WIELOPOLSKI, L
机构
[1] BROOKHAVEN NATL LAB,SUNY STONY BROOK,RADIAT THERAPY FACIL,BLDG 490,UPTON,NY 11973
[2] MED ACAD GDANSK,DEPT PHYS & BIOPHYS,PL-80211 GDANSK,POLAND
关键词
AGAR; ALANINE; ELECTRON PARAMAGNETIC RESONANCE; DOSIMETRY; PHANTOM;
D O I
10.1118/1.597130
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Calculations of the energy response of an agar-alanine phantom dosimeter (AAPD) to electrons in the energy range of 0.15-20 MeV together with experimental results at 16 MeV are presented. It is shown that the sensitivity of the EPR dosimeter (EPR signal/Gray) is independent of alanine crystal size and varies less than 2%, in the electron energy range indicated. Thus, the measured free radical density distribution may be used directly as an indication of the absorbed dose distribution in an irradiated phantom.
引用
收藏
页码:1453 / 1455
页数:3
相关论文
共 19 条
[11]  
HANSEN JW, 1987, RADIAT PROT DOSIM, V19, P43
[12]   ABSORBED DOSE AND CAVITY IONIZATION FOR HIGH-ENERGY ELECTRON BEAMS [J].
KESSARIS, ND .
RADIATION RESEARCH, 1970, 43 (02) :288-+
[13]   SPECTRA OF HIGH-ENERGY ELECTRON BEAMS IN WATER [J].
KESSARIS, ND .
RADIATION RESEARCH, 1970, 43 (02) :281-+
[14]   APPLICATIONS OF CAVITY THEORIES FOR ELECTRONS TO LIF DOSIMETERS [J].
PALIWAL, BR ;
ALMOND, PR .
PHYSICS IN MEDICINE AND BIOLOGY, 1975, 20 (04) :547-558
[15]   DOSIMETRY BY ELECTRON-SPIN-RESONANCE SPECTROSCOPY OF ALANINE [J].
REGULLA, DF ;
DEFFNER, U .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1982, 33 (11) :1101-+
[16]   CONTINUOUS 3-DIMENSIONAL RADIATION-DOSIMETRY IN TISSUE-EQUIVALENT PHANTOMS USING ELECTRON-PARAMAGNETIC RESONANCE IN L-ALPHA-ALANINE [J].
WIELOPOLSKI, L ;
MARYANSKI, M ;
CIESIELSKI, B ;
FORMAN, A ;
REINSTEIN, LE ;
MEEK, AG .
MEDICAL PHYSICS, 1987, 14 (04) :646-652
[17]   SEMIANALYTICAL EXPRESSIONS FOR (LBAR/P)(AIR)(MED) AND PREPL FOR ELECTRON-BEAMS [J].
WIELOPOLSKI, L ;
PAI, S ;
MLYN, M .
MEDICAL PHYSICS, 1991, 18 (03) :559-564
[18]  
1984, ICRU35 REP, P102
[19]  
1972, ICRU21 REP, P29