EX-VIVO EVALUATION OF FERROMAGNETISM, HEATING, AND ARTIFACTS PRODUCED BY HEART-VALVE PROSTHESES EXPOSED TO A 1.5-T MR SYSTEM

被引:43
作者
SHELLOCK, FG
MORISOLI, SM
机构
[1] ST JOHNS HOSP, SANTA MONICA, CA USA
[2] UNIV CALIF LOS ANGELES, SCH MED, DEPT RADIOL SCI, LOS ANGELES, CA USA
关键词
HEART; MR; PROSTHESES; SAFETY;
D O I
10.1002/jmri.1880040521
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Ex vivo testing techniques were used to determine the ferromagnetic qualities of, presence of heating in, and artifacts produced by 13 different heart valve prostheses exposed to a 1.5-T (64-MHz) magnetic resonance (MR) system. None of the heart valve prostheses showed a measurable deflection in the 1.5-T static magnetic field. Only minimal artifacts were produced during MR imaging with a fast spoiled GRASS (gradient-recalled acquisition in the steady state) pulse sequence. The largest temperature changes measured during a ''worst case'' MR imaging sequence (estimated average specific absorption rate, 2.5 W/kg; estimated spatial peak specific absorption rate, 7.6 W/kg) were +0.2-degrees-C with the implant imaged ''in air'' and +0.3-degrees-C with the implant imaged in normal saline. Therefore, MR procedures performed with a 1.5-T (64-MHz) MR system may be performed safely in patients with any of the 13 different heart valve prostheses evaluated in this study.