BLOOD-FLOW MEASUREMENT USING VARIABLE VELOCITY ENCODING IN THE RR INTERVAL

被引:35
作者
BUONOCORE, MH
机构
[1] Department of Radiology, UC Davis Medical Center, Sacramento, California
关键词
PULSE SEQUENCES; FLOW MEASUREMENT; PHASE CONTRAST; VARIABLE VELOCITY ENCODING;
D O I
10.1002/mrm.1910290611
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Velocity-encoded phase imaging using asynchronous gating requires input of a velocity encoding value to set the velocity sensitivity of the pulse sequence. The raw data interpolation and reconstruction scheme that the pulse sequence uses forces the encoding value to be constant throughout the RR interval. The sequence and the raw data interpolation scheme were modified to allow two velocity encodings during the RR interval. Two-hundred cm/s encoding was used in systole, and 30 cm/s in diastole. Changing the encoding in diastole significantly improved the accuracy and precision of ascending aorta flow measurements.
引用
收藏
页码:790 / 795
页数:6
相关论文
共 27 条
  • [1] BLOOD-FLOW IN THE CAROTID ARTERIES - QUANTIFICATION BY USING PHASE-SENSITIVE MR IMAGING
    BENDEL, P
    BUONOCORE, E
    BOCKISCH, A
    BESOZZI, MC
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1989, 152 (06) : 1307 - 1310
  • [2] BOGREN HG, 1992, WORKS IN PROGR S, pS1
  • [3] FACTORS INFLUENCING THE ACCURACY AND PRECISION OF VELOCITY-ENCODED PHASE IMAGING
    BUONOCORE, MH
    BOGREN, H
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 26 (01) : 141 - 154
  • [4] BUONOCORE MH, 1992, RADIOLOGY, V185, P160
  • [5] BUONOCORE MH, 1991, 10TH ANN M SOC MAGN, P803
  • [6] BUONOCORE MH, 1992, J MAGN RESON IMAGING, V2, P142
  • [7] BUONOCORE MH, 1991, 10TH ANN M SOC MAGN, P804
  • [8] POPLITEAL AND TIBIOPERONEAL ARTERIES - FEASIBILITY OF 2-DIMENSIONAL TIME-OF-FLIGHT MR ANGIOGRAPHY AND PHASE-VELOCITY MAPPING
    CAPUTO, GR
    MASUI, T
    GOODING, GAW
    CHANG, JM
    HIGGINS, CB
    [J]. RADIOLOGY, 1992, 182 (02) : 387 - 392
  • [9] RIGHT AND LEFT LUNG PERFUSION - INVITRO AND INVIVO VALIDATION WITH OBLIQUE-ANGLE, VELOCITY-ENCODED CINE MR IMAGING
    CAPUTO, GR
    KONDO, C
    MASUI, T
    GERACI, SJ
    FOSTER, E
    OSULLIVAN, MM
    HIGGINS, CB
    [J]. RADIOLOGY, 1991, 180 (03) : 693 - 698
  • [10] MR MEASUREMENT OF BLOOD-FLOW IN THE TRUE AND FALSE CHANNEL IN CHRONIC AORTIC DISSECTION
    CHANG, JM
    FRIESE, K
    CAPUTO, GR
    KONDO, C
    HIGGINS, CB
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1991, 15 (03) : 418 - 423