RADIOGRAPHIC DETECTION OF STRUT SEPARATIONS IN BJORK-SHILEY CONVEXOCONCAVE MITRAL-VALVES

被引:46
作者
ONEILL, WW [1 ]
CHANDLER, JG [1 ]
GORDON, RE [1 ]
BAKALYAR, DM [1 ]
ABOLFATHI, AH [1 ]
CASTELLANI, MD [1 ]
HIRSCH, JL [1 ]
WIETING, DW [1 ]
BASSETT, JS [1 ]
BEATTY, KC [1 ]
SOLTIS, MA [1 ]
TIMMIS, GC [1 ]
GRINES, CL [1 ]
机构
[1] SHILEY HEART VALVE RES CTR,IRVINE,CA
关键词
D O I
10.1056/NEJM199508173330703
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. The outlet struts of Bjork-Shiley convexo-concave heart valves can occasionally fracture. By December 31, 1994, 564 complete strut fractures had been reported to the manufacturer, approximately two thirds of which were fatal. There are no reliable diagnostic methods to detect waives that may be at risk for strut fracture. The outlet strut has two legs, and one leg often appears to break before the other, potentially permitting detection of the single-leg separation while the valve is still functionally intact. Methods. We used high-resolution cineradiography and defined valve profiles to evaluate 315 patients selected on the basis of their having mitral convexo-concave valves with an estimated fracture rate of 0.46 percent or higher per year. Two examinations were scheduled six months apart, with early reimaging performed when initial ratings were indeterminate. Results. Three patients had unsatisfactory studies, the most recent examinations in 277 patients were rated as apparently normal, 23 had findings considered minimally suspicious, and 1 had findings termed suspicious. The number of false negative results in this study group is unknown. Eleven cineradiograms were rated as showing probable or definite single-leg separations. All five ''definite'' ratings and five of the six ''probable'' ratings were confirmed by removal of the valves. One valve with a ''probable'' rating was intact. Two complete outlet-strut fractures occurred three and seven months after apparently normal radiographic examinations. Unsuspected new positive findings were not found at six months among 288 patients who completed the examination cycle. Conclusions. Cineradiographic imaging can detect some single-leg separations in mitral convexo-concave valves and may help the estimated 47,000 patients with these valves worldwide and their physicians decide about elective valve removal.
引用
收藏
页码:414 / 419
页数:6
相关论文
共 29 条
[1]  
BAKALYAR DM, 1994, CATHET CARDIOVASC DI, V32, P94
[2]   SHOULD PATIENTS WITH BJORK-SHILEY VALVES UNDERGO PROPHYLACTIC REPLACEMENT [J].
BIRKMEYER, JD ;
MARRIN, CAS ;
OCONNOR, GT .
LANCET, 1992, 340 (8818) :520-523
[3]  
BJORK VO, 1977, J THORAC CARDIOV SUR, V73, P563
[4]  
BJORK VO, 1970, J THORAC CARDIOV SUR, V60, P355
[5]  
Blackstone E H, 1992, J Heart Valve Dis, V1, P3
[6]  
BLACKSTONE EH, 1993, J HEART VALVE DIS, V2, P249
[7]   12-YEAR COMPARISON OF A BJORK-SHILEY MECHANICAL HEART-VALVE WITH PORCINE BIOPROSTHESES [J].
BLOOMFIELD, P ;
WHEATLEY, DJ ;
PRESCOTT, RJ ;
MILLER, HC .
NEW ENGLAND JOURNAL OF MEDICINE, 1991, 324 (09) :573-579
[8]   THROMBOSED BJORK-SHILEY MITRAL PROSTHESES [J].
COPANS, H ;
LAKIER, JB ;
KINSLEY, RH ;
COLSEN, PR ;
FRITZ, VU ;
BARLOW, JB .
CIRCULATION, 1980, 61 (01) :169-174
[9]   SINGLE-LEG STRUT FRACTURES IN EXPLANTED BJORK-SHILEY VALVES [J].
DEMOL, BA ;
KALLEWAARD, M ;
MCLELLAN, RB ;
VANHERWERDEN, LA ;
DEFAUW, JJ ;
VANDERGRAAF, Y .
LANCET, 1994, 343 (8888) :9-12
[10]   STRUT FRACTURE WITH BJORK-SHILEY 70-PERCENT CONVEXO-CONCAVE VALVE - AN INTERNATIONAL MULTIINSTITUTIONAL FOLLOW-UP-STUDY [J].
ERICSSON, A ;
LINDBLOM, D ;
SEMB, G ;
HUYSMANS, HA ;
THULIN, LI ;
SCULLY, HE ;
BENNETT, JG ;
OSTERMEYER, J ;
GRUNKEMEIER, GL .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1992, 6 (07) :339-346