FACTORS INFLUENCING TRACE ION ANALYSIS WITH PRECONCENTRATION BY ELECTROSTACKING

被引:42
作者
WOJTUSIK, MJ
HARROLD, MP
机构
[1] Dionex Corporation, Sunnyvale, CA 94088
关键词
D O I
10.1016/0021-9673(94)80268-8
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Capillary electrophoresis with indirect UV detection and electrostacking achieves low mu g/l detection limits for inorganic anions and cations in various low conductivity matrices. Since this technique requires electromigration injection, parameters such as analyte electrophoretic mobility, electroosmotic flow, and sample and operating buffer ionic strength influence the amount of material that is actually injected into the capillary. These parameters are not easily controlled and are affected by such factors as injection voltage and duration, operating buffer, and operating buffer pH. The effects of these factors on sensitivity and resolution are demonstrated. The use of a novel carrier ion, dimethyldiphenylphosphonium ion, improves sensitivity for cations. Techniques for optimizing trace level determinations are also discussed.
引用
收藏
页码:411 / 417
页数:7
相关论文
共 17 条
[1]   ELECTROOSMOTIC PROPERTIES AND PEAK BROADENING IN FIELD-AMPLIFIED CAPILLARY ELECTROPHORESIS [J].
CHIEN, RL ;
HELMER, JC .
ANALYTICAL CHEMISTRY, 1991, 63 (14) :1354-1361
[2]   SAMPLE STACKING OF AN EXTREMELY LARGE INJECTION VOLUME IN HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS [J].
CHIEN, RL ;
BURGI, DS .
ANALYTICAL CHEMISTRY, 1992, 64 (09) :1046-1050
[3]   MATHEMATICAL-MODELING OF FIELD-AMPLIFIED SAMPLE INJECTION IN HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS [J].
CHIEN, RL .
ANALYTICAL CHEMISTRY, 1991, 63 (24) :2866-2869
[4]   ON-COLUMN SAMPLE CONCENTRATION USING FIELD AMPLIFICATION IN CZE [J].
CHIEN, RL ;
BURGI, DS .
ANALYTICAL CHEMISTRY, 1992, 64 (08) :A489-A496
[5]   LARGE SAMPLE VOLUME PRESEPARATION FOR TRACE ANALYSIS IN ISOTACHOPHORESIS [J].
DOLNIK, V ;
DEML, M ;
BOCEK, P .
JOURNAL OF CHROMATOGRAPHY, 1985, 320 (01) :89-97
[6]   DETERMINATION OF SUBSTANCES AT LOW CONCENTRATIONS IN COMPLEX-MIXTURES BY ISOTACHOPHORESIS WITH COLUMN COUPLING [J].
EVERAERTS, FM ;
VERHEGGEN, TPEM ;
MIKKERS, FEP .
JOURNAL OF CHROMATOGRAPHY, 1979, 169 (FEB) :21-38
[7]  
Foret F, 1990, J MICROCOLUMN SEP, V2, P229
[8]   BIAS IN QUANTITATIVE CAPILLARY ZONE ELECTROPHORESIS CAUSED BY ELECTROKINETIC SAMPLE INJECTION [J].
HUANG, XH ;
GORDON, MJ ;
ZARE, RN .
ANALYTICAL CHEMISTRY, 1988, 60 (04) :375-377
[9]   OPTIMIZATION OF INJECTION TECHNIQUE IN CAPILLARY ION ELECTROPHORESIS FOR THE DETERMINATION OF TRACE-LEVEL ANIONS IN ENVIRONMENTAL-SAMPLES [J].
JACKSON, PE ;
HADDAD, PR .
JOURNAL OF CHROMATOGRAPHY, 1993, 640 (1-2) :481-487
[10]   OPTIMIZATION OF DETECTION SENSITIVITY IN THE CAPILLARY ELECTROPHORESIS OF INORGANIC ANIONS [J].
JANDIK, P ;
JONES, WR .
JOURNAL OF CHROMATOGRAPHY, 1991, 546 (1-2) :431-443