A MULTILEAF COLLIMATOR FIELD PRESCRIPTION PREPARATION SYSTEM FOR CONVENTIONAL RADIOTHERAPY

被引:13
作者
DU, MN
YU, CX
SYMONS, M
YAN, D
TAYLOR, R
MATTER, RC
GUSTAFSON, G
MARTINEZ, A
WONG, JW
机构
[1] Radiation Oncology, William Beaumont Hospital, Royal Oak
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1994年 / 30卷 / 03期
关键词
MULTILEAF COLLIMATOR; MLC; RADIATION FIELD SHAPING;
D O I
10.1016/0360-3016(92)90959-L
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The purpose of this work is to develop a prescription preparation system for efficient field shaping using a multileaf collimator that can be used in community settings as well as research institutions. The efficiency advantage of the computer-controlled multileaf collimator, over cerrobend blocks, to shape radiation fields has been shown in conformal treatments, which typically require complete volumetric computerized tomographic data for three-dimensional radiation treatment planning-a utility not readily available to the general community. As a result, most patients today are treated with conventional radiation therapy. Therefore, we believe that it is very important to fully use the same efficiency advantage of multileaf collimator as a block replacement in conventional practice. Methods and Material: The multileaf collimator prescription preparation systems developed by us acquires prescription images from different sources, including film scanner, and radiation treatment planning systems. The multileaf collimator angle and leaf positions are set from the desired field contour defined on the prescription image, by minimizing the area discrepancies. Interactive graphical tools include manual adjustment of collimator angle and leaf positions, and definition of portions of the field edges that require maximal conformation. Data files of the final leaf positions are transferred to the multileaf collimator controller via a dedicated communication link. Results: We have implemented the field prescription preparation system and a network model for integrating the multileaf collimator and other radiotherapy modalities for routine treatments. For routine plan evaluation, isodose contours measured with film in solid water phantom at prescription depth are overlaid on the prescription image. Preliminary study indicates that the efficiency advantage of the MLC over cerrobend blocks in conformal therapy also holds true for conventional treatments. Conclusion: Our model of computer-controlled prescription, evaluation, and treatment using multileaf collimators can be effectively implemented in both community settings and research institutions. The resultant increase in treatment efficiency and accuracy is now available for conventional radiotherapy.
引用
收藏
页码:707 / 714
页数:8
相关论文
共 5 条
[1]   COMPUTER-ASSISTED GENERATION OF MULTI-LEAF COLLIMATOR SETTINGS FOR CONFORMATION THERAPY [J].
HOUNSELL, AR ;
SHARROCK, PJ ;
MOORE, CJ ;
SHAW, AJ ;
WILKINSON, JM ;
WILLIAMS, PC .
BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (772) :321-326
[2]   THE DESIGN AND PERFORMANCE-CHARACTERISTICS OF A MULTILEAF COLLIMATOR [J].
JORDAN, TJ ;
WILLIAMS, PC .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (02) :231-251
[3]   THE USE OF A MULTI-LEAF COLLIMATOR FOR CONFORMAL RADIOTHERAPY OF CARCINOMAS OF THE PROSTATE AND NASOPHARYNX [J].
LOSASSO, T ;
CHUI, CS ;
KUTCHER, GJ ;
LEIBEL, SA ;
FUKS, Z ;
LING, CC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (02) :161-170
[4]   A MODEL FOR COMPUTER-CONTROLLED DELIVERY OF 3-D CONFORMAL TREATMENTS [J].
MAGERAS, GS ;
PODMANICZKY, KC ;
MOHAN, R .
MEDICAL PHYSICS, 1992, 19 (04) :945-953
[5]   DOSE DISTRIBUTIONS OF X-RAY FIELDS AS SHAPED WITH MULTILEAF COLLIMATORS [J].
ZHU, Y ;
BOYER, AL ;
DESOBRY, GE .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (01) :163-173