COMPARISON OF HOT CATHODE AND COLD-CATHODE IONIZATION GAUGES

被引:43
作者
PEACOCK, RN
PEACOCK, NT
HAUSCHULZ, DS
机构
[1] HPSDivision of MKSInstruments. Inc., Colorado, 80301, Boulder
来源
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A | 1991年 / 9卷 / 03期
关键词
D O I
10.1116/1.577439
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Ionization gauges are commonly used for pressure measurement from the lowest achievable pressures to 10(-2) Torr. Two types exist: In the hot cathode gauge (HCG) the source of the ionizing electrons is a thermionic cathode. In the cold cathode gauge (CCG) it is a circulating space charge current of electrons trapped in crossed electric and magnetic fields. Indications of both types of gauge are gas dependent. Relative sensitivities for different gases are not the same for different HCGs or CCGs, or for HCGs compared to CCGs. Both types of gauge are susceptible to contamination. The filament heating power of the HCG is sufficient to cause local outgassing of adsorbed material, and possibly large pressure errors. Each has certain spurious effects. The lower limit of the HCG is set by x-ray photocurrent and electron impact desorption. The CCG is subject to discontinuities in the current versus pressure characteristic. Starting of the discharge in a CCG may be delayed at ultrahigh vacuum (UHV). Pumping speeds of both types of gauge are of magnitude 10(-2) l/s, so that gauge pumping is usually negligible.
引用
收藏
页码:1977 / 1985
页数:9
相关论文
共 49 条
[21]  
NOTTINGHAM WB, 1962, 8TH 1961 T NAT VAC S, P4914
[22]  
OMAN RM, 1969, NASA CR1480 CONTR RE
[23]   SOME CHARACTERISTICS OF AN INVERTED MAGNETRON COLD-CATHODE IONIZATION GAUGE WITH DUAL FEEDTHROUGHS [J].
PEACOCK, NT ;
PEACOCK, RN .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1988, 6 (03) :1141-1144
[24]   THE INFLUENCE OF THE MAGNETIC-FIELD STRENGTH UPON THE CHARACTERISTIC CURVE OF A COLD-CATHODE IONIZATION GAUGE [J].
PEACOCK, NT ;
PEACOCK, RN .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1990, 8 (03) :2806-2809
[25]   SENSITIVITY VARIATION OF BAYARD-ALPERT GAUGES WITH AND WITHOUT CLOSED GRIDS FROM 10-4 TO 1 PA [J].
PEACOCK, RN ;
PEACOCK, NT .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1990, 8 (04) :3341-3344
[26]  
PEACOCK RN, 1990, 1990 P EUR C VAC TEC, P134
[27]   A new manometer for low gas pressures especially between 10(-3) and 10(-5) mm. [J].
Penning, FM .
PHYSICA, 1937, 4 :0071-0075
[28]   LONG-TERM BEHAVIOR OF IONIZATION GAUGES [J].
POULTER, KF .
VACUUM, 1981, 31 (03) :147-150
[29]   INSTABILITIES IN CROSSED-FIELD DISCHARGES AT LOW-PRESSURES [J].
REDHEAD, PA .
VACUUM, 1988, 38 (8-10) :901-906
[30]   MODULATED BAYARD-ALPERT GAUGE [J].
REDHEAD, PA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1960, 31 (03) :343-344