EXPERIMENTALLY DETERMINING A PARAMETRIC MODEL FOR THE POINT-SOURCE RESPONSE OF A GAMMA-CAMERA

被引:6
作者
HEBERT, T
MURPHY, P
MOORE, W
DHEKNE, R
WENDT, R
BLUST, M
机构
[1] BAYLOR COLL MED,DEPT RADIOL,HOUSTON,TX 77030
[2] ST LUKES EPISCOPAL HOSP,DEPT NUCL MED,HOUSTON,TX 77030
关键词
D O I
10.1109/23.256694
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this paper, we experimentally investigate the response of a gamma camera as a function of two parameters: depth of the source in water and remaining distance in air from the gamma camera. We demonstrate the linearity of the FWHM resolution measure versus these two parameters. We present a method for experimentally estimating the parameters of any camera response model from the experimental data. We use this approach to estimate the parameters of a 2-D Gaussian model for the camera point source response in a true 3-D formulation.
引用
收藏
页码:967 / 971
页数:5
相关论文
共 11 条
[1]   INVERSE MONTE-CARLO - A UNIFIED RECONSTRUCTION ALGORITHM FOR SPECT [J].
FLOYD, CE ;
JASZCZAK, RJ ;
COLEMAN, RE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1985, 32 (01) :779-785
[2]   PARAMETERIZATION OF THE SCATTER RESPONSE FUNCTION IN SPECT IMAGING USING MONTE-CARLO SIMULATION [J].
FREY, EC ;
TSUI, BMW .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (03) :1308-1315
[3]   SPATIAL PROPERTIES OF THE SCATTER RESPONSE FUNCTION IN SPECT [J].
FREY, EC ;
TSUI, BMW .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1991, 38 (02) :789-794
[4]   AN ATTENUATED PROJECTOR BACKPROJECTOR FOR ITERATIVE SPECT RECONSTRUCTION [J].
GULLBERG, GT ;
HUESMAN, RH ;
MALKO, JA ;
PELC, NJ ;
BUDINGER, TF .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (08) :799-816
[5]  
HEBERT TJ, 1992, IN PRESS IEEE T MAR
[6]  
MANBECK K, 1990, THESIS BROWN U
[7]   A PROJECTOR, BACK-PROJECTOR PAIR WHICH ACCOUNTS FOR THE 2-DIMENSIONAL DEPTH AND DISTANCE DEPENDENT BLURRING IN SPECT [J].
PENNEY, BC ;
KING, MA ;
KNESAUREK, K .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :681-686
[8]   MAXIMUM LIKELIHOOD APPROACH TO EMISSION IMAGE-RECONSTRUCTION FROM PROJECTIONS [J].
ROCKMORE, AJ ;
MACOVSKI, A .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1976, 23 (04) :1428-1432
[9]   IMPLEMENTATION OF SIMULTANEOUS ATTENUATION AND DETECTOR RESPONSE CORRECTION IN SPECT [J].
TSUI, BMW ;
HU, HB ;
GILLAND, DR ;
GULLBERG, GT .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :778-783
[10]  
ZEEBERG B R, 1987, Journal of Nuclear Medicine, V28, P662