PATIENT DOSE IN DUAL X-RAY ABSORPTIOMETRY

被引:110
作者
LEWIS, MK
BLAKE, GM
FOGELMAN, I
机构
[1] Department of Nuclear Medicine, Guy's Hospital, London, SE1 9RT, St. Thomas Street
关键词
BONE MINERAL MEASUREMENT; DUAL X-RAY ABSORPTIOMETRY; EFFECTIVE DOSE;
D O I
10.1007/BF02352255
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Dual x-ray absorptiometry (DXA) provides a convenient, non-invasive method of assessing skeletal bone mineral which is widely used for clinical studies. This report describes a study to estimate the effective dose of radiation (ICRP-60 (1990)) to a typical female patient from scans performed on three DXA scanners: the Hologic QDR-1000, QDR-1000/W and QDR-2000. The scans modes studied were: total body; anteroposterior (AP) lumbar spine; lateral lumbar spine; proximal femur; distal forearm. An ionization chamber and tissue-equivalent phantom were used to determine entrance surface dose and percentage depth-dose curves for each scan mode. Anatomical data from ICRP-23 (Reference Man) and a body section atlas were used to estimate the absorbed dose to each organ in the scan fields. Effective dose was estimated using the ICRP-60 tissue weighting factors and the fraction of each organ in the scan field. Results are summarized below. Figures for the effective dose are given both excluding and (in brackets) including the ovaries to cover the cases of postmenopausal and premenopausal women respectively.
引用
收藏
页码:11 / 15
页数:5
相关论文
共 21 条
  • [11] COMPARATIVE PERFORMANCE INVITRO AND INVIVO OF LUNAR DPX AND HOLOGIC QDR-1000 DUAL ENERGY X-RAY ABSORPTIOMETERS
    LASKEY, MA
    FLAXMAN, ME
    BARBER, RW
    TRAFFORD, S
    HAYBALL, MP
    LYTTLE, KD
    CRISP, AJ
    COMPSTON, JE
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (767) : 1023 - 1029
  • [12] PERFORMANCE EVALUATION OF A DUAL-ENERGY X-RAY BONE DENSITOMETER
    MAZESS, R
    COLLICK, B
    TREMPE, J
    BARDEN, H
    HANSON, J
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1989, 44 (03) : 228 - 232
  • [13] PURDIE D, 1992, CURRENT RES OSTEOPOR, V2, P84
  • [14] EFFECTIVE DOSE EQUIVALENT IN DUAL X-RAY ABSORPTIOMETRY
    PYE, DW
    HANNAN, WJ
    HESP, R
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1990, 63 (746) : 149 - 149
  • [15] RAWLINGS DJ, 1992, CURRENT RES OSTEOPOR, V2, P23
  • [16] REID DM, 1992, CURRENT RES OSTEOPOR, V2, P85
  • [17] SHRIMPTON PC, 1986, NRPBR200 NAT RAD PRO
  • [18] STEIGER P, 1991, OSTEOPOROSIS INT, V1, P190
  • [19] METHODS OF BONE-MINERAL MEASUREMENT
    TOTHILL, P
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (05) : 543 - 572
  • [20] COMPARISON OF DUAL-ENERGY X-RAY ABSORPTIOMETRY AND DUAL PHOTON-ABSORPTIOMETRY FOR BONE-MINERAL MEASUREMENTS OF THE LUMBAR SPINE
    WAHNER, HW
    DUNN, WL
    BROWN, ML
    RIGGS, BL
    [J]. MAYO CLINIC PROCEEDINGS, 1988, 63 (11) : 1075 - 1084